4K CAMERA HEAD
Report
- Report Number
- 8010047-2018-01651
- Event Type
- Malfunction
- Date Received
- August 27, 2018
- Date of Event
- August 1, 2018
- Report Date
- October 23, 2018
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- FET
- PMA / PMN Number
- K172817
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) CHECKED THE DEVICE HISTORY RECORD OF THE SUBJECT CH-S400-XZ-EB, THERE WAS NO IRREGULARITY FOUND. THE SUBJECT CH-S400-XZ-EB WAS NOT RETURNED TO OMSC FOR EVALUATION, OMSC COULD NOT EVALUATE THE SUBJECT CH-S400-XZ-EB. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED, HOWEVER THERE IS POSSIBILITY OF THIS PHENOMENON IS ATTRIBUTED TO FAILURE OF THE CAMERA CABLE OF THE SUBJECT CH-S400-XZ-EB BY THE FOLLOWING. CAMERA CABLE OF THE SUBJECT CH-S400-XZ-EB WAS BROKEN. OXIDATION WAS OCCURRED AT THE CONNECTOR OF THE SUBJECT CH-S400-XZ-EB. THE CH-S400-XZ-EB INSTRUCTION MANUAL STATES THE CORRESPONDING METHOD WHEN THERE IS AN ABNORMALITY FOR THE DEVICE. THERE WERE NO FURTHER DETAILS PROVIDED. IF SIGNIFICANT ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.
THE SUBJECT CH-S400-XZ-EB WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC). AS A RESULT OF INVESTIGATION, OMSC FOUND THE TRACE THAT THE CAMERA CABLE OF THE SUBJECT CH-S400-XZ-EB WAS PULLED STRONG. OMSC JUDGED THAT THE CAMERA CABLE OF THE SUBJECT CH-S400-XZ-EB WAS DISCONNECTED. OLYMPUS STATED THE APPROPRIATE HANDLING OF CH-S400-XZ-EB AND THE COUNTER MEASURES AGAINST ABNORMALITIES IN THE INSTRUCTION MANUAL OF CH-S400-XZ-EB AS THE FOLLOWING. DO NOT COIL THE CAMERA CABLE WITH A DIAMETER OF LESS THAN 20 CM. THE CAMERA CABLE MAY BE DAMAGED. BE CAREFUL NOT TO TWIST THE CAMERA CABLE WHEN IT IS COILED. THE CAMERA CABLE MAY BE DAMAGED. THE CABLE CAN BE COILED WITHOUT TWISTS BY PILING LOOPS THAT ARE MADE BY CROSSING THE CABLE IN FRONT AND BACK ALTERNATELY. NEVER EXCESSIVELY BEND, PULL, TWIST, COIL, SQUEEZE, OR CRUSH THE CAMERA CABLE, WHICH COULD DAMAGE THE CAMERA CABLE. NEVER EXCESSIVELY PULL THE CAMERA CABLE, BUT STRAIGHTEN IT GRADUALLY WHEN IT IS COILED. THE CAMERA CABLE COULD BE DAMAGED. DO NOT USE EXCESSIVE FORCE WHEN WIPING THE EXTERNAL SURFACES OF THE CAMERA CABLE. THE CAMERA CABLE COULD BE DAMAGED. NEVER ATTEMPT TO LIFT THE ENTIRE ASSEMBLY BY THE CAMERA CABLE WHILE THE CAMERA HEAD IS ATTACHED TO THE ENDOSCOPE. THE CAMERA CABLE COULD BE DAMAGED. NEVER USE A CLAMP OR FORCEPS TO ATTACH THE CAMERA CABLE TO ANOTHER OBJECT. THE CAMERA CABLE COULD BE DAMAGED. DO NOT PULL OUT THE VIDEO CONNECTOR BY PULLING THE CAMERA CABLE. EQUIPMENT DAMAGE MAY OCCUR. DO NOT ROTATE THE CAMERA HEAD TOO MANY TIMES IN THE SAME DIRECTION. THE CAMERA CABLE MAY BECOME TWISTED, CAUSING EQUIPMENT FAILURE. THEREFORE, OMSC CONCLUDED THAT THIS PHENOMENON IS ATTRIBUTED TO THE INAPPROPRIATE HANDLING BY THE USER. THERE WERE NO FURTHER DETAILS PROVIDED. IF SIGNIFICANT ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.
THE SUBJECT CH-S400-XZ-EB WAS NOT RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) YET. THE USER WILL NOT RETURN THE SUBJECT CH-S400-XZ-EB. THE CH-S400-XZ-EB INSTRUCTION MANUAL STATES THE CORRESPONDING METHOD WHEN THERE IS AN ABNORMALITY FOR THE DEVICE. THERE WERE NO FURTHER DETAILS PROVIDED. IF SIGNIFICANT ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.
DURING THE UNSPECIFIED PROCEDURE, THE IMAGE DISPLAYED ON THE UNSPECIFIED MONITOR WAS DISAPPEARED. THE USER REPLACED THE USED OTV-S400 WITH THE SIMILAR DEVICE AND COMPLETED THE PROCEDURE. (REFERENCE (B)(4))0 ACCORDING TO THE SUPPLEMENTAL INVESTIGATION, IT WAS FOUND THAT THIS PHENOMENON WAS NOT CAUSED BY OTV-S400, IT WAS CAUSED BY THE SUBJECT CH-S400-XZ-EB. THERE WERE NO REPORTS OF PATIENT INJURIES RELATED TO THIS INCIDENT.