FDA Adverse Event Malfunction Summary report: N

MTS ID TIPMASTER PIPETTOR

MDR report key: 781534 · Received November 10, 2006

Report

Report Number
1056600-2006-00280
Event Type
Malfunction
Date Received
November 10, 2006
Date of Event
May 20, 2006
Report Date
November 10, 2006
Manufacturer
MICRO TYPING SYSTEMS, INC.
Product Code
GJG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER REMOVED THE PIPETTOR FROM SERVICE AND PERFORMED MAINTENANCE USING THE TIPMASTER MAINTENANCE KIT. CLEANING, RE-GREASING AND HANGING THE O-RINGS BROUGHT THE DEVICE BACK TO EXPECTED OPERATION. THOUGH THE MOST LIKELY ROOT CAUSE OF THE EVENT IS USER ERROR IN MAINTAINING THE PIPETTOR, THE POSSIBILITY OF PIPETTE MALFUNCTION COULD NOT BE RULED OUT.

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT THE PIPETTOR WAS LEAKING. THIS COULD RESULT IN VARIABLE DELIVERY OF REAGENTS. DISPENSING VARIABLE VOLUMES OF REAGENTS/SAMPLES COULD CAUSE ERRONEOUS TEST RESULTS. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MTS ID TIPMASTER PIPETTOR MANUAL PIPETTOR GJG MICRO TYPING SYSTEMS, INC. MTS9640 NA

Patients

Seq Age Sex Outcome Treatment