MTS ID TIPMASTER PIPETTOR
MDR report key: 781534
·
Received November 10, 2006
Report
- Report Number
- 1056600-2006-00280
- Event Type
- Malfunction
- Date Received
- November 10, 2006
- Date of Event
- May 20, 2006
- Report Date
- November 10, 2006
- Manufacturer
- MICRO TYPING SYSTEMS, INC.
- Product Code
- GJG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER REMOVED THE PIPETTOR FROM SERVICE AND PERFORMED MAINTENANCE USING THE TIPMASTER MAINTENANCE KIT. CLEANING, RE-GREASING AND HANGING THE O-RINGS BROUGHT THE DEVICE BACK TO EXPECTED OPERATION. THOUGH THE MOST LIKELY ROOT CAUSE OF THE EVENT IS USER ERROR IN MAINTAINING THE PIPETTOR, THE POSSIBILITY OF PIPETTE MALFUNCTION COULD NOT BE RULED OUT.
Description of Event or Problem · 1
A CUSTOMER REPORTED THAT THE PIPETTOR WAS LEAKING. THIS COULD RESULT IN VARIABLE DELIVERY OF REAGENTS. DISPENSING VARIABLE VOLUMES OF REAGENTS/SAMPLES COULD CAUSE ERRONEOUS TEST RESULTS. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MTS ID TIPMASTER PIPETTOR | MANUAL PIPETTOR | GJG | MICRO TYPING SYSTEMS, INC. | MTS9640 | NA |