FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND HIGH PROFILE

MDR report key: 7806954 · Received August 22, 2018

Report

Report Number
1645337-2018-05121
Event Type
Injury
Date Received
August 22, 2018
Date of Event
July 19, 2018
Report Date
July 26, 2018
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001652
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION SUMMARY: UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITHOUT FLUID. WHITE MATERIAL WAS OBSERVED WITHIN THE DEVICE AND ON THE SHELL SURFACE. PRODUCT EVALUATION TEAM DISCOVERED TWO RENTS. RENT A MEASURING APPROXIMATELY 0.8 CM. MICROSCOPIC EXAMINATION OF THE EDGES OF THE RENT A REVEALED PARALLEL STRIATIONS; AND RENT B MEASURING APPROXIMATELY < 0.1 CM WITHIN A CREASE LOCATED BOTH ON THE POSTERIOR ASPECT. NO OTHER ANOMALIES WERE DISCOVERED. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND IN THE DEVICE. MICROSCOPIC EXAMINATION REVEALED PARALLEL STRIATIONS. THIS TYPE OF STRIATIONS IS MORE CONCLUSIVE TO SHARP INSTRUMENT DAMAGE RATHER THAN SHELL FAILURE DUE TO WEAR. ALL THE IMPLANTS ARE 100% INSPECTED BEFORE LEAVING THE FACILITY. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. CONCOMITANT PRODUCTS: RIGHT MENTOR SMOOTH ROUND HIGH PROFILE 380CC SALINE BREAST IMPLANT, CATALOG NUMBER 3503380, LOT NUMBER 5921304, SERIAL NUMBER (B)(4). MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) HISPANIC FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH MENTOR SMOOTH ROUND HIGH PROFILE 380CC SALINE BREAST IMPLANTS WHICH THE LEFT SIDE DEFLATED AFTER IMPLANTATION. AS A RESULT, PATIENT UNDERWENT BILATERAL REMOVAL AND REPLACEMENT AS FOLLOW: LEFT REPLACED WITH CATALOG NUMBER 3504501BC. SERIAL NUMBER (B)(4), AND RIGHT REPLACED WITH CATALOG NUMBER 3504501BC, SERIAL NUMBER (B)(4) ON (B)(6) 2018.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
646171 MENTOR SMOOTH ROUND HIGH PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 5921304 00081317001652

Patients

Seq Age Sex Outcome Treatment
1 47 YR Required Intervention