FDA Adverse Event Malfunction Summary report: N

ENDOWRIST

MDR report key: 7805492 · Received August 22, 2018

Report

Report Number
2955842-2018-10475
Event Type
Malfunction
Date Received
August 22, 2018
Date of Event
July 20, 2018
Report Date
July 24, 2018
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
PMA / PMN Number
K112263
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE TIP COVER ACCESSORY INVOLVED WITH THIS COMPLAINT AND COMPLETED THE DEVICE EVALUATION. FAILURE ANALYSIS CONFIRMED THE CUSTOMER REPORTED FAILURE. THE DISTAL END OF THE TIP COVER EXHIBITED LOCALIZED MELTING WHICH IS INDICATIVE OF ARCING. A HOLE APPROXIMATELY 0.13¿ X 0.07¿ WAS VISIBLE AT THE BODY OF THE TIP COVER. MATERIAL WAS MISSING. ADDITIONALLY, THE MCS INSTRUMENT ASSOCIATED WITH THIS EVENT WAS NOT RETURNED FOR EVALUATION. THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: IT WAS ALLEGED THAT AFTER COMPLETION OF DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CUSTOMER NOTED "MELTING" ON THE MCS TIP COVER ACCESSORY. ALTHOUGH, NO PATIENT HARM WAS REPORTED, THE MELTING IS INDICATIVE OF ARCING. RECURRENCE OF THE REPORTED FAILURE MODE COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER COMPLETION OF A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE CUSTOMER INSPECTED THE MONOPOLAR CURVED SCISSORS (MCS) TIP COVER ACCESSORY AND FOUND IT WAS "MELTED". THERE WAS NO REPORT OF FRAGMENTS FALLING INSIDE THE PATIENT, PATIENT HARM, ADVERSE OUTCOME OR INJURY. INTUITIVE SURGICAL, INC. (ISI) MADE AN ATTEMPT TO FOLLOW UP AND OBTAIN ADDITIONAL INFORMATION, HOWEVER, NO FURTHER DETAILS HAVE BEEN RECEIVED AS OF DATE OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
645655 ENDOWRIST TIP COVER ACCESSORY NAY INTUITIVE SURGICAL, INC 400180-14 M11180215

Patients

Seq Age Sex Outcome Treatment
1