FDA Adverse Event Malfunction Summary report: N

VIDAS® RUB IGG II

MDR report key: 7803044 · Received August 21, 2018

Report

Report Number
3002769706-2018-00154
Event Type
Malfunction
Date Received
August 21, 2018
Report Date
November 6, 2018
Manufacturer
BIOMERIEUX SA
Product Code
LFX
PMA / PMN Number
K080766
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

A CUSTOMER IN SWITZERLAND REPORTED A FALSE POSITIVE RESULT FOR EQA (EXTERNAL QUALITY ASSESSMENT ) SAMPLES WHEN TESTING WITH VIDAS® RUB IGG II (REF. 30221) LOT 1006264820. THE CONCERNED EQA SAMPLE IS 2018-2R1 FROM SRK ORGANIZATION (SWISS EQA PROVIDER). AN INTERNAL INVESTIGATION WAS PERFORMED. THE ANALYSIS OF THE BATCH HISTORY RECORDS OF VIDAS RUB IGG II REFERENCE 30221 BATCH 1006264820 / 190125-0 SHOWED NO ANOMALY DURING THE MANUFACTURING, CONTROL AND PACKAGING PROCESSES. A COMPLAINT TRENDING ANALYSIS CONFIRMED THERE IS NO OTHER COMPLAINT RECORDED ON VIDAS RUB IGG II REFERENCE 30221 BATCH 1006264820 / 190125-0 CONCERNING EQA. STUDY OF INTERNAL SAMPLES CONTROL CHARTS INCLUDED: FIVE INTERNAL SAMPLES WITH DIFFERENT CONCENTRATIONS (TARGET : 13.3UI/ML, 11.5 UI/ML, 9.87 UI/ML AND 4.07 UI/ML), FIVE BATCHES INCLUDING THE BATCH MENTIONED BY THE CUSTOMER. VIDAS RUB IGG II REFERENCE 30221 BATCH 1006264820 / 190125-0 IS IN THE TREND ANALYSIS COMPARED TO THE OTHER BATCHES. TEST PERFORMED: FIVE INTERNAL SAMPLES (TARGET : 13.3UI/ML, 11.5 UI/ML, 9.87 UI/ML AND 4.07 UI/ML) WERE TESTED ON THE VIDAS RUB IGG II REFERENCE 30221 BATCH 1006264820 / 190125-0. ALL THE SAMPLES WERE WITHIN THEIR ACCEPTABLE RANGE. THERE WAS NO OBSERVATION OF ANY EVOLUTION, COMPARED TO THE RESULTS OBTAINED BEFORE THE RELEASE OF VIDAS RUB IGG II REFERENCE 30221 BATCH 1006264820 / 190125-0. THE CUSTOMER SAMPLE, EQA SRK 2018-2R1, WAS TESTED USING VIDAS RUB IGG II REFERENCE 30221 BATCH 1006264820 / 190125-0 AND THE RESULT WAS FOUND POSITIVE (42UI/ML). THE CUSTOMER RESULT WAS REPRODUCED. IT WAS NOT POSSIBLE TO PURSUE FURTHER TESTS AS THERE WAS NO REMAINING VOLUME. OTHER EQA ANALYSIS: AS BIOMÉRIEUX SUBSCRIBES TO EXTERNAL QUALITY ASSESSMENT PROGRAMS, AN ANALYSIS OF THE RESULTS OBTAINED ON CTCB PROGRAM (FRENCH EQA PROVIDER PROGRAM) IN 2018 FOR VIDAS RUB IGG II WAS PERFORMED. THIS ANALYSIS SHOWED THAT RESULTS FOR TWO POSITIVE CONTROLS WERE COMPLIANT TO THE PEER GROUP RESULTS AND REACHED THE EXPECTATIONS. CONCLUSION: TESTING REPRODUCED THE POSITIVE RESULT WHEN TESTING SRK 2018-R1 WITH VIDAS RUB IGG II REFERENCE 30221 BATCH 1006264820 / 190125-0. NEVERTHELESS, ACCORDING TO SRK REPORT, THIS POSITIVE RESULT OBTAINED FOR EQA SRK 2018-2R1USING VIDAS RUB IGG II REFERENCE 30221 BATCH 1006264820 / 190125-0 IS CONSIDERED AS CONFORMING. THERE IS NO INFORMATION IN HAND REGARDING THE CONTROL COMPOSITION, OR THE METHOD USED TO DETERMINE THE EQUIVOCAL STATUS OF THIS QUALITY CONTROL MATERIAL. MOREOVER, QUALITY CONTROL SAMPLES ARE NOT IDENTICAL TO PATIENT SAMPLES, QUALITY CONTROL MANUFACTURING PROCESS CAN AFFECT SAMPLE MATRIX AND CONTROL RESULTS. ACCORDING TO THE DATA MENTIONED ABOVE, THE BATCH VIDAS RUB IGG II REFERENCE 30221 BATCH 1006264820 / 190125-0 IS WITHIN THE EXPECTED PERFORMANCES.

Description of Event or Problem · 1

A CUSTOMER IN (B)(6) NOTIFIED BIOMÉRIEUX OF FALSE POSITIVE RESULTS WHEN TESTING A QUALITY CONTROL SAMPLE WITH VIDAS® RUB IGG II (REF. 30221), LOT 1006264820. THE CUSTOMER PERFORMED A ROUND ROBIN TEST AS PART OF AN EXTERNAL QC TRIAL FROM SRK. THE RESULT IS EXPECTED TO BE BORDERLINE/EQUIVOCAL. THE TECHNIQUES USED TO TEST THE SAMPLES WERE: ABBOTT RUBELLA, ARCHITECT; BIOMÉRIEUX RUBELLA; DIASORIN; ROCHE RUBELLA; SIEMENS RUBELLA. THE CUSTOMER TESTED THE SAMPLE TWICE, AND BOTH RESULTS WERE POSITIVE AT 42 IU/ML AND 46 IU/ML. AS THERE IS NO PATIENT ASSOCIATED WITH THIS QUALITY CONTROL STRAIN, THERE IS NO ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
644968 VIDAS® RUB IGG II VIDAS® RUB IGG II LFX BIOMERIEUX SA 1006264820

Patients

Seq Age Sex Outcome Treatment
1