FDA Adverse Event Malfunction Summary report: N

HYGENIC

MDR report key: 7800496 · Received August 21, 2018

Report

Report Number
2416455-2018-00002
Event Type
Malfunction
Date Received
August 21, 2018
Date of Event
June 18, 2018
Report Date
August 29, 2018
Manufacturer
COLTENE WHALEDENT INC.
Product Code
EEF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EG
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 0

THE ISSUE WAS REPORTED TO BE, B2 CLAMP AFTER PUTTING IT ON THE TOOTH IN MINUTES JUMPED IN MY FACE AND INJURED MY PATIENT. UPON FURTHER INVESTIGATION IT WAS FOUND THAT THE CLAMP BROKE IN THE PATIENT'S MOUTH AND CAUSED A SMALL CUT. PRODUCT WAS NOT RETURNED, HOWEVER SEVERAL PICTURES WERE PROVIDED. THE PICTURES DO NOT SHOW THE CLAMP DATE CODE NOR WAS A BATCH NUMBER PROVIDED, SO IT IS UNKNOWN WHEN THE CLAMP WAS MANUFACTURED. FROM THE PICTURES, THE CLAMP APPEARS TO BE TARNISHED AND HAS RESIDUE ON THE SURFACE. THE SURFACE OF THE BREAK SPECIFICALLY LOOKS BLACK. THIS SUGGESTS THAT THE CLEANING SOLUTION USED IS HARMFUL TO THE METAL AND THAT THE CLAMPS ARE NOT GETTING PROPERLY RINSED AFTER CLEANING. IT WAS DETERMINED THAT THE CUSTOMER USES UNODENT ENZYMATIC ULTRASONIC ECONOMY REFILL TO CLEAN THE CLAMP. THIS CLEANING SOLUTION CONTAINS DIDECYLDIMETHYL AMMONIUM CHLORIDE, WHICH IS CAUSTIC. THE RESIDUE ON THE CLAMP SUGGESTS IT WAS NOT BEING RINSED PROPERLY. IMPROPER RINSING AFTER USING A CAUSTIC CLEANING SOLUTION LEAVES THE CAUSTIC SOLUTION ON THE SURFACE WHICH LEADS TO DEGRADATION OF THE METAL. THE BLACK COLOR ON THE SURFACE OF THE BREAK INDICATES THE PRESENCE OF CAUSTIC STRESS CORROSION (CSC), WHICH IS THE MOST PROBABLE CAUSE OF THE CLAMP BREAKAGE. TO PREVENT CSC, THE CLAMPS NEED TO BE RINSED WITH WATER AFTER CLEANING DE-IONIZED IS PREFERABLE AS TAP WATER MAY CONTAIN HARSH ALKALINE CHEMICALS (CHLORIDES) ADDED TO KILL BACTERIA. INSUFFICIENT RINSING AFTER USING CAUSTIC CHEMICALS IN THE CLEANING PROCESS PRIOR TO AUTOCLAVE STERILIZATION WILL SIGNIFICANTLY INCREASE THE CHANCE OF PREMATURE BREAKAGE AS THE HEAT FROM THE AUTOCLAVE WILL SPEED UP THE DEGRADATION PROCESS. THIS IS THE ONLY COMPLAINT FOR THIS PART NUMBER IN THE PAST 12 MONTHS. RISK ASSESSMENT: PROBABILITY: IMPROBABLE. SEVERITY: SERIOUS. RISK: MINIMAL RESIDUAL RISK. CAPA REQUIRED: NO.

Description of Event or Problem · 1

AFTER PUTTING THE DENTAL CLAMP ON TOOTH THE CLAMP POPPED OUT OF MOUTH AND INJURED PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
641184 HYGENIC DENTAL CLAMP EEF COLTENE WHALEDENT INC. H01049

Patients

Seq Age Sex Outcome Treatment
1