FDA Adverse Event Malfunction Summary report: N

SERVO-U

MDR report key: 7797256 · Received August 20, 2018

Report

Report Number
8010042-2018-00431
Event Type
Malfunction
Date Received
August 20, 2018
Report Date
January 23, 2019
Manufacturer
MAQUET CRITICAL CARE AB
Product Code
CBK
PMA / PMN Number
K151814
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC. 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4). THE FAULT CONDITION COULD NOT BE REPRODUCED OR CONFIRMED, WHEN THE VENTILATOR WAS INVESTIGATED ON SITE BY OUR FIELD SERVICE ENGINEER. AS A PREVENTIVE MEASURE THE AIR GAS MODULE WAS REPLACED. DESPITE SEVERAL REMINDERS THE AIR GAS MODULE WAS NOT RETURNED FOR OUR INVESTIGATION EVALUATION OF THE RECEIVED DEVICE LOGS SHOWS THAT THE MATCHING EVENT ON JULY 9 DURING THE TIME PERIOD 02:07 TO 02:44. THE ROOT CAUSE WHY THE VENTILATOR GENERATED ALARMS FOR HIGH AIR SUPPLY PRESSURE HAS NOT BEEN DETERMINED. BUT DUE TO THE REPORTED PART AND THAT NO NEW EVENT HAS BEEN REPORTED WE STILL CONCLUDE THAT MOST LIKELY CASE WAS A DEFECTIVE AIR GAS MODULE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE VENTILATOR GENERATED AN ALARM FOR HIGH AIR SUPPLY PRESSURE. PATIENT INVOLVEMENT IS UNKNOWN. IMPORTER REF. #: (B)(4). MANUFACTURER REF. #: (B)(4).

Description of Event or Problem · 0

IMPORTER REF. #: (B)(4). MANUFACTURER REF. #: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
637331 SERVO-U VENTILATOR, CONTINUOUS, FACILITY USE CBK MAQUET CRITICAL CARE AB

Patients

Seq Age Sex Outcome Treatment
1