GUTTA-PERCHA PLUGGER
Report
- Report Number
- 8031010-2018-00083
- Event Type
- Malfunction
- Date Received
- August 17, 2018
- Report Date
- August 28, 2018
- Manufacturer
- DENTSPLY MAILLEFER
- Product Code
- EKR
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- DENTIST
Narratives
RETURNED PLUGGER HAS ONE OF TIP (#2) WHICH IS ACTUALLY BROKEN IN THE SECOND DEPTH RING. NO MATERIAL DEFECT WAS FOUND DURING ANALYSIS OF THE RUPTURE PATTERN. THE BATCH NUMBER IS UNKNOWN, DHR CANNOT BE REVIEWED. ROOT CAUSES ARE NOT IDENTIFIED. THIS KIND OF EVENT IS TRACKED AND WE MONITOR THE TREND.
WHILE NO SERIOUS INJURY RESULTED IN THIS EVENT, THERE HAS BEEN A PREVIOUS REPORT RECEIVED WHERE THIS MALFUNCTION RESULTED IN A SERIOUS INJURY. THEREFORE, IT MUST BE PRESUMED THAT RECURRENCE OF THIS MALFUNCTION COULD POSSIBLY CAUSE OR CONTRIBUTE TO A SERIOUS INJURY OR REQUIRE MEDICAL OR SURGICAL INTERVENTION TO PRECLUDE SUCH. AS SUCH, THIS EVENT IS REPORTABLE PER 21CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH RESULTS ARE NOT AVAILABLE AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
IN THIS EVENT IT WAS REPORTED THAT A PLUGGER BROKE DURING USE; NO INJURY RESULTED AND THE SEPARATED PIECE WAS RETRIEVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 636418 | GUTTA-PERCHA PLUGGER | PLUGGER, ROOT CANAL, ENDODONTIC | EKR | DENTSPLY MAILLEFER | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |