FDA Adverse Event Injury Summary report: N

MENTOR SILTEX ROUND MODERATE PROFILE

MDR report key: 7791547 · Received August 16, 2018

Report

Report Number
1645337-2018-04997
Event Type
Injury
Date Received
August 16, 2018
Date of Event
July 20, 2018
Report Date
July 27, 2018
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001775
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

SINCE NO LOT NUMBER WAS PROVIDED, NO DEVICE HISTORY RECORD (DHR) REVIEW COULD BE PERFORMED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITHOUT FLUID. YELLOW MATERIAL WAS OBSERVED WITHIN THE DEVICE AND ON THE SHELL SURFACE. PRODUCT EVALUATION DISCOVERED A RENT MEASURING APPROXIMATELY 0.1 CM WITHIN AN AREA OF SILTEX CRACKING ON THE POSTERIOR ASPECT. NO OTHER ANOMALIES WERE DISCOVERED. SINCE THE LOT NUMBER WAS NOT PROVIDED BY THE CUSTOMER, THE DEVICE HISTORY RECORD (DHR) COULD NOT BE REVIEWED. IF LOT NUMBER IS LATER PROVIDED, THE DHR WILL BE REVIEWED. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE OR THE ETIOLOGY OF THE YELLOW MATERIAL FOUND ON THE DEVICE. ALL THE IMPLANTS ARE 100% INSPECTED BEFORE LEAVING THE FACILITY. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. SILTEX CRACKING IS DEFINED AS A MATERIAL SEPARATION OCCURRING IN THE SILTEX LAYER AND CAN EVENTUALLY PROGRESS THROUGH THE SHELL OF SILTEX DEVICES. MENTOR BELIEVES THAT SILTEX CRACKING IS ULTIMATELY A RESULT OF HIGH STRESSES CAUSED BY ACUTE FOLDS (I.E., SEVERE ¿V¿ OR TIGHT FOLDS) IN THE RTV SHELL OF SILTEX BREAST IMPLANTS. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. (B)(4).

Additional Manufacturer Narrative · 0

ON 9/18/2018, THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A DEVICE HISTORY RECORD REVIEW IS IN PROGRESS. ONCE COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: MENTOR SILTEX ROUND MODERATE PROFILE 275CC SALINE, CATALOG: 3542640, LOT: 252752, SN: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) -YEAR OLD CAUCASIAN FEMALE PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH MENTOR SILTEX ROUND MODERATE PROFILE 275CC SALINE BREAST PROSTHESES. DEFLATION OF THE LEFT BREAST PROSTHESIS WAS CONFIRMED. AS A RESULT, THE PATIENT WAS SCHEDULED FOR REMOVAL ON (B)(6) 2018.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
630306 MENTOR SILTEX ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 00081317001775

Patients

Seq Age Sex Outcome Treatment
1 45 YR Required Intervention