Description of Event or Problem · 1
WHEN THE MIST TENT IS SET-UP PER MFR'S RECOMMENDATIONS A STATIC PROBLEM IS PRESENT. WITH THE UNIT PROPERLY SET-UP AND RUNNING, IF ONE HOLDS THE THREE OUTLET HOSES IN ONE HAND AND TOUCHES THE EQUIPMENT CHASIS WITH THE OTHER HAND ONE WILL RECEIVE A STATIC SHOCK. THIS WAS DUPLICATED BY THE MFR ON SITE AT HOSP. EQUIPMENT WAS NEVER USED ON PT. MFR'S ATTITUDE WAS THAT THIS PROBLEM WAS UNIQUE TO THIS INSTITUTION ONLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ALLIED HEALTHCARE PRODUCTS, INC. | O2/MIST TENT | BYL | ALLIED HEALTHCARE PRODUCTS, INC. | 15100 CAM 4 | NA |