ELECTA WASH SET 225 ML BOWL
Report
- Report Number
- 1718850-2006-00028
- Event Type
- Malfunction
- Date Received
- August 30, 2006
- Date of Event
- June 20, 2006
- Report Date
- June 23, 2006
- Manufacturer
- SORIN GRP ITALIA S.R.I, A SORIN GRP CO., DISTRIBUTED BY COBE CARDIOVASCULAR, INC
- Product Code
- CAC
- PMA / PMN Number
- k020647
- Removal / Correction Number
- PENDING
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
H6) THE PROBLEM PRODUCT INVOLVED IN THIS REPORT WAS NOT RETURNED FOR EVALUATION. THE CONLUSION IS BASED ON EVALUATION OF WASH SETS IN A BRACKET THAT INCLUDED THIS LOT. THE EVALUATION REVEALED THAT IF THE BOWL CANNOT BE INSERTED IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE, THERE IS A RISK THAT THE BOWL'S ROTATING SEAL AREA COULD DETERIORATE DURING CENTRIFUGING, RELEASING PARTICULATE INTO THE PROCESSED BLOOD. THE LOADING OF WASH SETS FROM THIS BRACKET OF PRODUCT CAN BE MORE DIFFICULT DUE TO A MANUFACTURING PROCESS ERROR THAT RESULTED IN A SLIGHT DEFORMATION IN THE MOUNTING RING ON THE BOWL BOTTOM. THE PROCESS ERROR HAS SUBSEQUENTLY BEEN CORRECTED BY THE MANUFACTURER. BASED ON THE EVENT DESCRIBED IN G4 ABOVE, SORIN GROUP ITALIA COMPLETED A RISK ASSESSEMENT THAT LEAD TO A FIELD NOTIFICATION. THE FIELD NOTIFICATION IN THE U.S. WAS CONDUCTED BY COBE CARDIOVASCULAR, INC., THE DISTRIBUTOR OF THIS DEVICE. THE Z-NUMBER WILL BE PROVIDED WHEN IT IS AVAILABLE FROM THE LOCAL FDA OFFICE.
A NOISE WAS HEARD FROM 2 AUTOLOGOUS TRANSFUSION WASH SET BOWLS FROM THIS LOT DURING USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ELECTA WASH SET 225 ML BOWL | APPARATUS, AUTOTRANSFUSION | CAC | SORIN GRP ITALIA S.R.I, A SORIN GRP CO., DISTRIBUTED BY COBE CARDIOVASCULAR, INC | NA | 0601180100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN |