FDA Adverse Event Malfunction Summary report: N

HEARTWARE VENTRICULAR ASSIST SYSTEM-BATTERY

MDR report key: 7789853 · Received August 16, 2018

Report

Report Number
3007042319-2018-03838
Event Type
Malfunction
Date Received
August 16, 2018
Date of Event
July 30, 2018
Report Date
January 16, 2019
Manufacturer
HEARTWARE, INC.
Product Code
DSQ
UDI-DI
00888707000369
PMA / PMN Number
P100047
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

PRODUCT EVENT SUMMARY: THE BATTERY WAS RETURNED FOR EVALUATION. VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICE IN RELATION TO THE REPORTED EVENT. FAILURE ANALYSIS OF THE BATTERY REVEALED THAT THE DEVICE PASSED VISUAL EXAMINATION AND FUNCTIONAL TESTING. THE BATTERY WAS ABLE TO ADEQUATELY PROVIDE POWER TO A TEST CONTROLLER. IN ADDITION, THE BATTERY WAS CHARGING WHEN CONNECTED TO A BATTERY CHARGER. A REVIEW OF THE BATTERY'S INTERNAL LOG REVEALED THAT A CELL PAIR, AT ONE POINT IN TIME, PASSED THE VOLTAGE THRESHOLD. WHEN THIS OCCURS, THE BATTERY WILL NOT CHARGE UNTIL THE VOLTAGE DECREASES BELOW THE THRESHOLD. HOWEVER, THE BATTERY WAS RECEIVED WITH NO FLAGS ENABLED. IF THE BATTERY REMAINS CONNECTED TO THE BATTERY CHARGER WITH A CELL-OVER-VOLTAGE FLAG, THE BATTERY CHARGER STATUS INDICATOR WILL FLASH RED AFTER 8 HOURS DUE TO A CHARGE TIME-OUT. AS A RESULT, THE REPORTED BATTERY NOT CHARGING EVENT WAS CONFIRMED. HOWEVER, THE REPORTED BATTERY NO POWER AND FLASHING YELLOW EVENTS COULD NOT BE CONFIRMED. BASED ON THE AVAILABLE INFORMATION, A POSSIBLE ROOT CAUSE OF THE BATTERY NOT CHARGING EVENT CAN BE ATTRIBUTED TO AN OVERVOLTAGE FAULT DETECTED BY THE ANALOG FRONT END (AFE) INTEGRATED CIRCUIT. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION AND DESTINATION THERAPY IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BATTERY WAS NOT CHARGING. THE INDICATOR LIGHT ON THE BATTERY CHARGER WAS FLASHING YELLOW. THE BATTERY WAS TESTED ON THE OTHER CONTROLLER PORT AND NO POWER WAS PROVIDED. THE BATTERY WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
632028 HEARTWARE VENTRICULAR ASSIST SYSTEM-BATTERY VENTRICULAR (ASSISST) BYPASS DSQ HEARTWARE, INC. 1650DE 00888707000369

Patients

Seq Age Sex Outcome Treatment
1 VAD| VAD