FDA Adverse Event Malfunction Summary report: N

HARMONIC SCALPEL CURVED SHEAR

MDR report key: 778620 · Received November 3, 2006

Report

Report Number
1527736-2006-05742
Event Type
Malfunction
Date Received
November 3, 2006
Date of Event
August 23, 2006
Report Date
August 29, 2006
Manufacturer
ETHICON ENDO SURGERY, INC.(CINCINNATI)
Product Code
LFL
PMA / PMN Number
k002981
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE LCSC5 DEVICE WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. DUE TO THE DAMAGE TO THE BLADE, IT WAS CONFIRMED THAT THE DEVICE WAS NON-FUNCTIONAL. THE DEVICE WAS TESTED WITH A GENERATOR AND AN ERROR CODE 5 WAS NOTED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. FOR THIS REASON THE FOLLOWING STATEMENTS WERE INCLUDED IN THE INSTRUCTIONS FOR USE: "AVOID ACCIDENTAL CONTACT WITH ALL METAL OR PLASTIC INSTRUMENTS OR OBJECTS WHEN THE INSTRUMENT IS ACTIVATED. CONTACT WITH STAPLES, CLIPS OR OTHER INSTRUMENTS WHILE THE INSTRUMENT IS ACTIVATED MAY RESULT IN CRACKED OR BROKEN BLADES, WHICH MAY BE IDENTIFIED BY GENERATOR SOLID TONE OR INSTRUMENT ERROR." EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED AT THIS PROCESS.

Description of Event or Problem · 1

IT WAS REPORTEDTHAT DURING A LAP GASTRIC BYPASS R-EN-Y PROCEDURE, LCSC5 TESTED CORRECTLY AND THEN FAILED TO WORK AFTER A FEW MINUTES OF OPERATING. CASE COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE. NO PT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HARMONIC SCALPEL CURVED SHEAR GEI LFL ETHICON ENDO SURGERY, INC.(CINCINNATI) NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN HANDPIECE| GENERATOR