FDA Adverse Event Malfunction Summary report: N

25 GA HIGH SPEED VITRECTOMY CUTTER

MDR report key: 7784119 · Received August 15, 2018

Report

Report Number
0001920664-2018-00098
Event Type
Malfunction
Date Received
August 15, 2018
Date of Event
July 3, 2018
Report Date
July 23, 2018
Manufacturer
BAUSCH + LOMB
Product Code
HQC
UDI-DI
20757770057928
PMA / PMN Number
K101325
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

REFER TO CAPA (B)(4). CAPA (B)(4) WAS CLOSED ON 06 JULY 2018 BUT WAS STILL OPEN WHEN LOT V8936 WAS MANUFACTURED IN MARCH 2017. PER CAPA (B)(4) AND (B)(4) THE VITRECTOMY TUBESET USED ON THE STELLARIS PC PACKS AND STANDALONE POUCHED VITRECTOMY CUTTERS WAS REPLACED WITH A NEW VITRECTOMY TUBESET. THE NEW VITRECTOMY TUBESET PROVIDES A GREATER OPERATING RANGE BETWEEN THE PRESSURE REQUIRED TO CLOSE THE VITRECTOMY CUTTER INNER NEEDLE AND THE PRESSURE REQUIRED TO OPEN THE VITRECTOMY CUTTER INNER NEEDLE. THE VITRECTOMY TUBESET ENHANCEMENT WAS IMPLEMENTED ON THE IMPACTED PC PACKS AND POUCHED ACCESSORIES. THE FIRST LOT OF BL5625 THAT WAS BUILT WITH THE VITRECTOMY TUBESET ENHANCEMENT WAS LOT #W2161 IN JULY 2018. ANY BL5625 PRODUCT WITH LOT NO. W2161 OR HIGHER WILL INCLUDE THE VITRECTOMY TUBESET ENHANCEMENT. REPORT 2 OF 2 SEE 1920664-2018-00097.

Additional Manufacturer Narrative · 1

THE STERILIZATION AND LOT HISTORY RECORDS WERE REVIEWED AND FOUND TO BE ACCEPTABLE. THE PRODUCT COMPLAINT CANNOT BE CONFIRMED, THERE WAS NO PRODUCT RETURNED FOR EVALUATION. FURTHER INVESTIGATION UNDERWAY. REPORT 2 OF 2 SEE: 1920664-2018-00097.

Description of Event or Problem · 1

THE USER FACILITY IN (B)(6) REPORTED THE VITRECTOMY CUTTERS WOULD NOT CUT WHILE ASPIRATION CONTINUED. THE CUTTER WAS REPLACED WITH ANOTHER, AND THE SURGERY WAS COMPLETED WITH NO PATIENT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
628562 25 GA HIGH SPEED VITRECTOMY CUTTER UNIT, PACOFRAGMENTATION HQC BAUSCH + LOMB BL5625 V8936 20757770057928

Patients

Seq Age Sex Outcome Treatment
1