FDA Adverse Event Death Summary report: N

GEM PREMIER 5000

MDR report key: 7784100 · Received August 15, 2018

Report

Report Number
1217183-2018-00007
Event Type
Death
Date Received
August 15, 2018
Date of Event
June 29, 2018
Report Date
August 15, 2018
Manufacturer
INSTRUMENTATION LABORATORY
Product Code
CHL
UDI-DI
08426950870227
PMA / PMN Number
K173403
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INSTRUMENTATION LABORATORY (IL) CONDUCTED AN INVESTIGATION THAT INCLUDED A REVIEW OF THE DATA FILES FROM THE CUSTOMER'S GEM PREMIER 5000 INSTRUMENTS. THE DATA REVIEW SHOWED THAT THE DEVICES WERE OPERATING WITHIN SPECIFICATIONS. THERE WAS NO SYSTEM OR SENSOR MALFUNCTION DURING ANALYSIS OF THE PATIENT'S SAMPLES. FOR THE SAMPLES IN QUESTION, ALL RECOVERIES WERE REPORTED AS INCALCULABLE, EXCEPT CO-OX AND ITS FRACTIONS. IT SHOULD BE NOTED THAT THE RESULTS WERE FLAGGED, INCLUDING GLUCOSE, AS "INCALCULABLE", WHICH THE OPERATOR'S MANUAL INDICATES AS "REQUIRED MEASUREMENT CRITERIA WERE NOT MET DURING SAMPLE ANALYSIS". THE INVESTIGATION CONCLUDED THAT THE INSTRUMENTS WERE OPERATING WITHIN SPECIFICATION AND, THEREFORE, NO REMEDIAL ACTION IS REQUIRED.

Description of Event or Problem · 1

A COMPLAINT WAS RECEIVED INDICATING A DEATH IN A PATIENT WHO DEVELOPED HYPOGLYCEMIA FOLLOWING A MEDICATION ERROR WHERE INSULIN WAS ADMINISTERED INSTEAD OF HEPARIN BECAUSE OF AN INCORRECTLY LABELED SYRINGE PUMP. THE COMPLAINT STATES THAT ARTERIAL SAMPLES WERE TESTED ON MULTIPLE GEM PREMIER 5000 ANALYZERS ON (B)(6) 2018. THE ELECTROCHEMICAL RESULTS WERE NOT REPORTED AND THERE WERE NO ERROR MESSAGES. AS A RESULT, THE COMPLAINT ALLEGES A DELAY IN OBTAINING GLUCOSE RESULTS, WHICH IF AVAILABLE SOONER COULD HAVE ASSISTED THE CLINICAL TEAM WITH THE PATIENT MANAGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
628911 GEM PREMIER 5000 CLINICAL CHEMISTRY INSTRUMENT CHL INSTRUMENTATION LABORATORY 2262 08426950870227

Patients

Seq Age Sex Outcome Treatment
1 94 YR Death