FDA Adverse Event Malfunction Summary report: N

LUMENIS ONE

MDR report key: 7775834 · Received August 12, 2018

Report

Report Number
3004135191-2018-00107
Event Type
Malfunction
Date Received
August 12, 2018
Report Date
August 12, 2018
Manufacturer
LUMENIS, LTD.
Product Code
GEX
PMA / PMN Number
K060448
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

A FOREIGN USER FACILITY REPORTED THAT A COOLING SYSTEM LEAK OCCURRED IN A LUMENIS ONE LASER. NO REPORT OF RELATED INJURY WAS RECEIVED. LUMENIS INVESTIGATED THE REPORTED COMPLAINT. A LUMENIS TECHNICAL EXPERT COMPLETED AN EXAMINATION OF THE SUBJECT DEVICE IDENTIFYING A CRACK IN A COOLANT TUBE FITTING TO BE THE CAUSE OF THE EVENT REPORT. LUMENIS IS REPORTING THIS COMPLAINT SINCE IT IS SIMILAR TO AN EVENT IN WHICH THE COOLING SYSTEM LEAK RESULTED IN COMBUSTION OF THE HIGH VOLTAGE POWER SUPPLY. THE ORIGINAL MDR WAS MDR 3004135191-2012-00063-(B)(4).

Description of Event or Problem · 1

A FOREIGN USER FACILITY REPORTED THAT A COOLING SYSTEM LEAK OCCURRED IN A LUMENIS ONE LASER. NO REPORT OF RELATED INJURY WAS RECEIVED.