FDA Adverse Event Malfunction Summary report: N

HANDPIECE MICS

MDR report key: 7772482 · Received August 10, 2018

Report

Report Number
3005985723-2018-00477
Event Type
Malfunction
Date Received
August 10, 2018
Date of Event
July 31, 2018
Report Date
September 21, 2018
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
UDI-DI
00848486030193
PMA / PMN Number
K142530
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

FOLLOW-UP #1 AND FINAL REPORT SUBMITTED TO UPDATE SECTIONS BASED ON THE RESULTS OF INVESTIGATION. REPORTED ISSUE HANDPIECE ROTATION RELEASE ON TOP OF POWER SAW, CAME OFF WHEN DR. TRIED TO TURN HANDLE. CASE TYPE: TKA. PRODUCT HISTORY REVIEW DEVICE HISTORY RECORDS INDICATE 25 DEVICES WERE MANUFACTURED UNDER LOT K076W AND ALL 25 INCLUDING 4200566 WERE ACCEPTED INTO FINAL STOCK ON 4/14/16. COMPLAINT HISTORY REVIEW A REVIEW OF COMPLAINTS RELATED TO P/N 209063, S/N (B)(4) IN PRODEX LOT K076W SHOWS NO ADDITIONAL COMPLAINT(S) RELATED TO THE FAILURE IN THIS INVESTIGATION. VISUAL INSPECTION VISUAL INSPECTION REVEALED THAT THE HANDLE PIVOT LATCH DISLOCATED FROM THE HANDPIECE. DIMENSIONAL INSPECTION DIMENSIONAL INSPECTION WAS NOT COMPLETED AS VISUAL AND FUNCTIONAL INSPECTION CLEARLY SHOWS THE FAILURE OF THE DEVICE. MATERIAL ANALYSIS MATERIAL ANALYSIS WAS NOT COMPLETED AS VISUAL AND FUNCTIONAL INSPECTION CLEARLY SHOWS THE FAILURE OF THE DEVICE. FUNCTIONAL INSPECTION FUNCTIONAL INSPECTION REVEALED THAT THE HANDPIECE HANDLE COULD NOT BECOME UNLOCKED BECAUSE THE LATCH TO UNLOCK THAT FUNCTION HAD DISLOCATED. CONCLUSION IT WAS CONFIRMED THAT THE PIN ATTACHING THE HANDLE PIVOT LATCH TO THE MICS BECAME DISLOCATED.

Description of Event or Problem · 0

HANDPIECE ROTATION RELEASE ON TOP OF POWER SAW, CAME OFF WHEN DR. TRIED TO TURN HANDLE. CASE TYPE: TKA.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

HANDPIECE ROTATION RELEASE ON TOP OF POWER SAW, CAME OFF WHEN DR. TRIED TO TURN HANDLE. CASE TYPE: TKA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
613924 HANDPIECE MICS STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 42020316/ 4200566 00848486030193

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization