Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS INCIDENT HAS BEEN RECOGNIZED AS USER ERROR BY VARIAN MEDICAL SYSTEMS. LABELING PLACED ON THE COUCH, AS WELL AS IN THE USER AND SAFETY MANUAL, PROVIDE INSTRUCTION AND HAZARD WARNINGS (WITH RESPECT TO HAND PLACEMENT), WHICH HELPS TO ENSURE THE SAFE OPERATION OF THE EQUIPMENT.

Description of Event or Problem · 1

FOUR (4) THERAPIST WERE SETTING UP A PT FOR TREATMENT. THEY RELEASED THE LONGITUDINAL LOCKS, AND MANUALLY MOVED THE PT INTO THE TREATMENT FIELD. ONE OF THE THERAPISTS WAS ON THE KLYSTRON SIDE OF THE TABLE AT THE FOOT OF THE PT. AS THE LONGITUDINAL CARRIAGE WAS MOVED INTO THE FIELD, THE THERAPIST'S LEFT RING FINGER WAS CAUGHT BETWEEN THE LATERAL CARRIAGE ASSEMBLY AND THE STRETCHER ASSEMBLY. THE THERAPIST NEEDED SURGERY ON HER FINGER AND WAS HOSPITALIZED DUE TO THIS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EXACT COUCH PT POSITIONING TABLE JAI VARIAN MEDICAL SYSTEMS EXACT COUCH NA

Patients

Seq Age Sex Outcome Treatment
1 40 YR Hospitalization| O