WHITE STAR HIGH SPEED VIT CUTTER
Report
- Report Number
- 3006695864-2018-01651
- Event Type
- Malfunction
- Date Received
- August 9, 2018
- Date of Event
- July 13, 2018
- Report Date
- November 2, 2020
- Manufacturer
- JOHNSON & JOHNSON SURGICAL VISION, INC
- Product Code
- MLZ
- UDI-DI
- 05050474534117
- PMA / PMN Number
- K081681
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE AVAILABLE FOR EVALUATION? YES. RETURNED TO MANUFACTURER ON: 08/10/2018. DEVICE RETURNED TO MANUFACTURER? YES. DEVICE EVALUATION: THE DISPOSABLE VITRECTOMY CUTTER WAS RETURNED WITHIN ITS ORIGINAL PACKAGING CONFIRMING THE REPORTED LOT NUMBER OF 60087423. A VISUAL INSPECTION OF THE RETURNED DEVICE SHOWED NEITHER THE INNER OR OUTER PACKAGE WAS OPENED. FAILURE MODE OF MISLABELED SAMPLE WAS CONFIRMED AS THERE IS A 20GA VIT CUTTER LABELED AS 25GA. THE REPORTED MISLABELING ISSUE WAS CONFIRMED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE NR-0148919 AND CAPA-010215.
(B)(6). ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
A SURGERY CENTER REPORTED MISLABELING ON TWO DISPOSABLE VITRECTOMY CUTTERS. A BRIEF DESCRIPTION FROM THE SURGERY CENTER INDICATED THEY ORDERED SIZE 25-GAUGE DISPOSABLE VITRECTOMY CUTTERS BUT UPON OPENING THE PACKAGE, THE ACTUAL VITRECTOMY CUTTER HAD SIZE 20-GAUGE ENGRAVED ON THE DEVICE. NO INDICATION OF PATIENT CONTACT. THIS IS 1 OF 2 REPORTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 612200 | WHITE STAR HIGH SPEED VIT CUTTER | DISPOSABLE VITRECTOMY CUTTER | MLZ | JOHNSON & JOHNSON SURGICAL VISION, INC | NGP0025 | 60087423 | 05050474534117 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |