VENA SEAL CLOSURE SYSTEM
Report
- Report Number
- 9612164-2018-01983
- Event Type
- Injury
- Date Received
- August 8, 2018
- Date of Event
- July 9, 2018
- Report Date
- October 3, 2018
- Manufacturer
- MEDTRONIC IRELAND
- Product Code
- PJQ
- PMA / PMN Number
- P140018
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IMAGE REVIEW: THREE SONOGRAPHIC IMAGES WERE RECEIVED FOR EVALUATION. THE THREE IMAGES INCLUDE THE SFJ, COMMON FEMORAL VEIN, AND FEMORAL ARTERY. IMAGE 1 IS OF THE LEFT COMMON FEMORAL VEIN AND THE SFJ IN A LONGITUDINAL PLANE. IMAGE 2 IS OF THE LEFT COMMON FEMORAL VEIN AND THE SFJ IN A TRANSVERSE PLANE. IMAGE 3 IS OF THE LEFT COMMON FEMORAL VEIN AND THE SFJ IN A LONGITUDINAL PLANE. A MASS OF THROMBOSIS CAN BE SEEN IN ALL THREE IMAGES. BASED ON THE QUALITY AND CLARITY OF THE IMAGES IT IS NOT POSSIBLE TO POSITIVELY DETERMINE IF THE THROMBOSIS MASS INCLUDES ADHESIVE. NO COMPONENTS FROM THE VENASEAL CLOSURE SYSTEM KIT WERE RECEIVED FOR EVALUATION. NO PHOTOGRAPHIC IMAGES OF THE PATIENT¿S SYMPTOMS WERE RECEIVED FOR EVALUATION. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT A PATIENT WAS TREATED WITH VENA SEAL AS PER THE IFU IN THE GREAT SAPHENOUS VEIN(GSV). APPROXIMATELY A 30CM LENGTH WAS TREATED WITH 1 CM OF CYANOACRYLATE. THE CATHETER TIP LOCATION WAS 5CM CAUDAL TO THE SAPHENOFEMORAL JUNCTION (SFJ) PRIOR TO INITIAL DELIVERY OF THE ADHESIVE AS PER THE IFU. THE PATIENT RETURNED FOR A FOLLOW UP STUDY AND IT WAS FOUND THAT GLUE HAD EXTENDED CLOSE TO THE COMMON FEMORAL VEIN. THE PATIENT WAS PRESCRIBED BLOOD THINNERS AS A RESULT OF THE GLUE EXTENSION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 602047 | VENA SEAL CLOSURE SYSTEM | AGENT, OCCLUDING, VASCULAR, PERMANENT | PJQ | MEDTRONIC IRELAND |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |