FDA Adverse Event Injury Summary report: N

EP-WORKMATE¿ SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 56 CHANNEL

MDR report key: 7761730 · Received August 8, 2018

Report

Report Number
2184149-2018-00122
Event Type
Injury
Date Received
August 8, 2018
Date of Event
July 11, 2018
Report Date
October 2, 2018
Manufacturer
ST. JUDE MEDICAL, INC.
Product Code
DQK
UDI-DI
05415067003126
PMA / PMN Number
K151911
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED STARTUP FAILURE AND CANCELLED PROCEDURE COULD NOT BE CONCLUSIVELY DETERMINED. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. THE REVIEW DETERMINED THE PROCESS WAS PERFORMED AND COMPLETED IN ACCORDANCE WITH ABBOTT SPECIFICATIONS AND PROCEDURES.

Description of Event or Problem · 1

DURING AN ATRIOVENTRICULAR NODAL REENTRY TACHYCARDIA PROCEDURE A CANCELLATION OCCURRED DUE TO SIGNAL LOSS. DURING THE PROCEDURE ECG AND INTRACARDIAC SIGNALS WERE NOT AVAILABLE. THE ISSUE COULD NOT BE RESOLVED AND THE PROCEDURE WAS CANCELLED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
604436 EP-WORKMATE¿ SCU-SIGNAL CONDITIONING UNIT (AMPLIFIER), 56 CHANNEL PROGRAMMABLE DIAGNOSTIC COMPUTER DQK ST. JUDE MEDICAL, INC. 09-1462-0056 05415067003126

Patients

Seq Age Sex Outcome Treatment
1 Other