FDA Adverse Event
Malfunction
Summary report: N
AIRLIFE¿ INFANT HEATED WIRE CIRCUIT
MDR report key: 7760097
·
Received August 7, 2018
Report
- Report Number
- 8030673-2018-00006
- Event Type
- Malfunction
- Date Received
- August 7, 2018
- Date of Event
- June 24, 2018
- Report Date
- August 15, 2018
- Manufacturer
- VYAIRE MEDICAL
- Product Code
- OGL
- PMA / PMN Number
- ENFORCEMENT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 0
THE DEVICE SAMPLE WAS NOT AVAILABLE FOR INVESTIGATION BY OUR QUALITY TEAM. HOWEVER, THE QUALITY TEAM IDENTIFIED THAT A DEFECTIVE CRIMPING COULD CAUSE BAD ELECTRICAL CONNECTION OF HEATING WIRE TO THE POWER SUPPLY. THIS COULD GENERATE RAIN OUT IN THE CIRCUIT. IF SET UP TO CRIMPING EQUIPMENT IS NOT PROPERLY DONE, THE CRIMPING HEIGHT CAN BE OUT OF SPEC.
Additional Manufacturer Narrative · 1
AT THIS TIME, VYAIRE HAS NOT RECEIVED THE SAMPLE FOR INVESTIGATION. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
THE R.N. REPORTED: I FIND THIS TUBING TO HAVE MORE "RAIN-OUT" WATER IN TUBING. RESPIRATORY THERAPIST HAS TO CONSTANTLY COME AND REMOVE WATER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 596937 | AIRLIFE¿ INFANT HEATED WIRE CIRCUIT | OXYGEN ADMINISTRATION KIT | OGL | VYAIRE MEDICAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |