UNIPOC DATA MANAGEMENT SYSTEM
Report
- Report Number
- 3002637618-2018-00092
- Event Type
- Malfunction
- Date Received
- August 7, 2018
- Date of Event
- July 12, 2018
- Report Date
- August 8, 2018
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS
- Product Code
- JQP
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ADDITIONAL INFORMATION WAS ADDED TO EVENT DESCRIPTION.
A NON-SIEMENS COAGULATION MONITORING INSTRUMENT CONNECTED TO UNIPOC BY THE MIDDLEWARE DNOTE (SIEMENS-DISTRIBUTED) DISPLAYED AN ACT+ RESULT OF ">1005" ON THE INSTRUMENT DISPLAY (AN OUT-OF-RANGE HIGH VALUE), HOWEVER THE INSTRUMENT TRANSMITTED THE RESULT "0" TO DNOTE AND AN INDICATOR THAT THE RESULT WAS "NORMAL". DNOTE THEN PASSED THE RESULT TO UNIPOC, WHICH DISPLAYED THE "0" RESULT TO THE CUSTOMER. A MESSAGE "ERROR: OUT OF RANGE-HI" WAS PRESENT IN THE NTE (COMMENTS) FIELD. IT HAS BEEN CONFIRMED THAT DNOTE OPERATED AS INTENDED. THERE WAS NO REPORT OF INJURY DUE TO THIS EVENT.
DNOTE AND UNIPOC ARE RECEIVING AND DISPLAYING THE DATA AS SENT BY THE INSTRUMENT. SIEMENS HAS CONTACTED THE INSTRUMENT MANUFACTURER TO REPORT THIS ISSUE.
A NON-SIEMENS COAGULATION MONITORING INSTRUMENT CONNECTED TO UNIPOC BY THE MIDDLEWARE DNOTE (SIEMENS-DISTRIBUTED) DISPLAYED AN ACT+ RESULT OF ">1005" ON THE INSTRUMENT DISPLAY (AN OUT-OF-RANGE HIGH VALUE), HOWEVER THE INSTRUMENT TRANSMITTED THE RESULT "0" TO DNOTE AND AN INDICATOR THAT THE RESULT WAS "NORMAL". DNOTE THEN PASSED THE RESULT TO UNIPOC, WHICH DISPLAYED THE "0" RESULT TO THE CUSTOMER. THERE WAS NO REPORT OF INJURY DUE TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 598962 | UNIPOC DATA MANAGEMENT SYSTEM | UNIPOC & DNOTE | JQP | SIEMENS HEALTHCARE DIAGNOSTICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |