FDA Adverse Event Malfunction Summary report: N

HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0

MDR report key: 7758128 · Received August 7, 2018

Report

Report Number
3007042319-2018-03525
Event Type
Malfunction
Date Received
August 7, 2018
Date of Event
July 18, 2018
Report Date
November 20, 2018
Manufacturer
HEARTWARE, INC.
Product Code
DSQ
PMA / PMN Number
P100047
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

CORRECTED INFORMATION: MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

PRODUCT EVENT SUMMARY: FOUR BATTERIES, ONE CONTROLLER AC ADAPTER, ONE CONTROLLER DC ADAPTER, AND ONE CONTROLLER WERE NOT RETURNED FOR EVALUATION. LOG FILE ANALYSIS WAS NOT CONDUCTED SINCE LOG FILES WERE NOT AVAILABLE. AS A RESULT, THE REPORTED EVENT COULD NOT BE CONFIRMED. A POWER SOURCE LUBRICATION PROCEDURE WAS PERFORMED TO MITIGATE THE REPORTED CONDITIONS AND THE POWER SOURCES REMAIN IN USE. APPLICABLE RISK DOCUMENTATION AND EXPERIENCE WITH EVENTS OF SIMILAR CIRCUMSTANCES WERE CONSIDERED; A POSSIBLE ROOT CAUSE OF THE REPORTED PREMATURE POWER SWITCHING EVENTS CAN BE ATTRIBUTED TO COMMUNICATION ERRORS AND/OR MOMENTARY DISCONNECTIONS BETWEEN THE CONTROLLER AND POWER SOURCES. ADDITIONAL PRODUCTS: BATTERY / BAT597520 FDA METHOD CODE(S): 4114, 4112 FDA RESULTS CODE(S): 3221 FDA CONCLUSION CODE(S): 67 BATTERY / BAT597516 FDA METHOD CODE(S): 4114, 4112 FDA RESULTS CODE(S): 3221 FDA CONCLUSION CODE(S): 67 BATTERY / BAT597521 FDA METHOD CODE(S): 4114, 4112 FDA RESULTS CODE(S): 3221 FDA CONCLUSION CODE(S): 67 BATTERY / BAT597519 H6 FDA METHOD CODE(S): 4114, 4112 FDA RESULTS CODE(S): 3221 FDA CONCLUSION CODE(S): 67 CONTROLLER AC ADAPTER / CAC200591 H6 FDA METHOD CODE(S): 4114, 4112 FDA RESULTS CODE(S): 3221 FDA CONCLUSION CODE(S): 67 CONTROLLER DC ADAPTER / CDC004835 FDA METHOD CODE(S): 4114, 4112 FDA RESULTS CODE(S): 3221 FDA CONCLUSION CODE(S): 67 MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION AND DESTINATION THERAPY IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. OTHER DEVICES INVOLVED IN THIS EVENT: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY/ (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2019-03-31 UDI #:(B)(4) MFG DATE: 2018-03-27 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY/ (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2019-03-31 UDI #: (B)(4) MFG DATE: 2018-03-27 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿BATTERY/ (B)(4) / MODEL #: 1650DE / EXPIRATION DATE: 2019-03-31 UDI #: (B)(4) MFG DATE: 2018-03-28 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY/ (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2019-03-31 UDI #: (B)(4) MFG DATE: 2018-03-27 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿CONTROLLER AC ADAPTOR/ (B)(4) / MODEL #: 1430DE (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ CONTROLLER DC ADAPTOR / (B)(4)/ MODEL #: 1440/ UDI #: (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCE PREMATURE POWER SWITCHING ON THE CONTROLLER AND MULTIPLE POWER SOURCES. THE CONTROLLER AND ALL ASSOCIATED POWER SOURCES REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
601657 HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0 VENTRICULAR (ASSISST) BYPASS DSQ HEARTWARE, INC. 1420-CONTROLLER

Patients

Seq Age Sex Outcome Treatment
1 MCS VAD