BENT HOHMANN RETRACTOR
Report
- Report Number
- 3005985723-2018-00463
- Event Type
- Malfunction
- Date Received
- August 7, 2018
- Date of Event
- July 9, 2018
- Report Date
- October 9, 2018
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- UDI-DI
- 00848486031909
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 AND FINAL REPORT SUBMITTED TO UPDATE SECTIONS BASED ON THE RESULTS OF INVESTIGATION. PER THE MPS, ALL THE RETRACTORS WERE DAMAGED. PRODUCT NOT RETURNED. REFER TO (B)(4). A REVIEW OF THE DEVICE HISTORY RECORDS INDICATE 199 DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK ON (B)(6) 2018. NO NON-CONFORMANCES WERE IDENTIFIED DURING INSPECTION. A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 210191, LOT NUMBER 37899 SHOWS NO ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. SYSTEM READY FOR CLINICAL USE. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD."
RETRACTOR BADLY DAMAGED DUE TO SAW HITTING IT. PER THE MPS, ALL THE RETRACTORS WERE DAMAGED AND RIGID. CASE TYPE: (B)(6).
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
RETRACTOR BADLY DAMAGED DUE TO SAW HITTING IT. PER THE MPS, ALL THE RETRACTORS WERE DAMAGED AND RIGID. CASE TYPE: TKA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 601019 | BENT HOHMANN RETRACTOR | STEREOTAXIC DEVICE, ROBOTICS | OLO | MAKO SURGICAL CORP. | 37899 | 00848486031909 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |