SITTER ELITE
Report
- Report Number
- 2020362-2018-00067
- Event Type
- Malfunction
- Date Received
- August 7, 2018
- Report Date
- July 10, 2018
- Manufacturer
- POSEY PRODUCTS LLC
- Product Code
- PJO
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS DID NOT CONFIRM THE ALLEGED FAILURE OF THE ALARM ALERTING WHEN IT WASN'T SUPPOSED TO. THE DEVICE WAS RECEIVED IN WITH SCRATCHES TO THE EXTERIOR HOUSING AND MISSING BATTERY DOOR. THE BATTERY COMPARTMENT ALSO SHOWED SIGNS OF BATTERY CORROSION WITH WHITE POWDER RESIDUE OBSERVED ON THE BATTERY CONTACT, SPRINGS AND COMPARTMENT WALL. DURING FUNCTIONAL TESTING, THE UNIT DID NOT SOUND WHEN IN USE WITH THE PULL MAGNET AND SENSOR PAD. UPON OPENING THE DEVICE, BATTERY CORROSION WAS FOUND ON DIFFERENT PARTS OF THE PRINTED CIRCUIT BOARD. THIS LOSS OF FUNCTIONALITY WOULD RESULT IN THE ALARM'S FAILURE TO ALERT THE CAREGIVER OF A PATIENT EXIT AND COULD CONTRIBUTE TO A PATIENT INCIDENT. BEING THAT THE DEVICE IS OVER FIVE YEARS OLD, IT IS LIKELY EXPECTED NORMAL WEAR AND TEAR CONTRIBUTED TO THE DEFECT. AT THIS TIME, THERE IS NO EVIDENCE THAT A MANUFACTURING NON-CONFORMITY CONTRIBUTED TO THE REPORTED COMPLAINT AND THE INSTRUCTIONS FOR USE (IFU) WERE REVIEWED AND DETERMINED TO PROVIDE ADEQUATE INSTRUCTIONS AND WARNING FOR THE SAFE AND EFFECTIVE USE OF THE DEVICE. THE IFU STATES "TO REDUCE THE RISK OF SERIOUS INJURY OR DEATH, TEST THE ALARM AND SENSOR FOR PROPER OPERATION PRIOR TO PUTTING IN SERVICE WITH A PATIENT, AND EACH TIME BEFORE LEAVING THE PATIENT UNATTENDED. IF THE ALARM AND/OR SENSOR DO NOT FUNCTION PROPERLY, REMOVE THE ALARM AND SENSOR FROM SERVICE AND REPLACE THEM WITH A PROPERLY FUNCTIONING ALARM AND/OR SENSOR. DO NOT USE THE ALARM, SENSOR OR MAGNET IF IT DOES NOT ACTIVATE EACH TIME WEIGHT IS REMOVED FROM THE SENSOR, THE CHAIR BELT SENSOR IS UNFASTENED, OR MAGNET IS REMOVED FROM FACE PLATE." THEREFORE, NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE NECESSARY. ALL COMPLAINTS ARE TRENDED AND REVIEWED BY MANAGEMENT ON A MONTHLY BASIS. AS PART OF THIS MONTHLY REVIEW, ANY EXCURSION ABOVE THE CONTROL LIMITS FOR THIS FAILURE MODE WILL BE ASSESSED, DOCUMENTED AND ACTED UPON AS WARRANT. (B)(4).
CUSTOMER REPORTED ALARM GOES OFF WHEN WEIGHT IS ON A NEW SENSOR. INVESTIGATION FOUND THE UNIT HAS NO SOUND. THE DATE THE ISSUE WAS DISCOVERED IS UNKNOWN AND NO PATIENT INCIDENT OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 596590 | SITTER ELITE | FALL PREVENTION ALARM/SENSOR COMBINATION ATTACHED OR UNATTACHED | PJO | POSEY PRODUCTS LLC | 8345 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |