FDA Adverse Event
Malfunction
Summary report: N
SINGLE FIBER OPTIC CABLE, 10FT, PINK
MDR report key: 7755286
·
Received August 6, 2018
Report
- Report Number
- 3011394215-2018-00005
- Event Type
- Malfunction
- Date Received
- August 6, 2018
- Date of Event
- July 11, 2018
- Report Date
- July 11, 2018
- Manufacturer
- INVUITY, INC
- Product Code
- FST
- UDI-DI
- 00816728020202
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
DURING EVALUATION, IT WAS NOTED THAT BROKEN FIBERS WERE PRESENT THROUGHOUT THE CABLE. ADDITIONALLY, THE CABLE EXHIBITED A CONCENTRATED NUMBER OF BROKEN FIBERS WHICH INDICATES MISHANDLING OF THE CABLE. THE CONCENTRATION OF BROKEN FIBERS LED TO AN ELEVATED TEMPERATURE. INVUITY'S IFU CLEARLY STATES TO HANDLE THE CABLES WITH CARE AND NOT TO COIL THE CABLES TIGHTER THAN SIX (6) INCHES.
Description of Event or Problem · 1
DURING A HIP REPLACEMENT PROCEDURE, IT WAS OBSERVED THAT THE DRAPE HAD A BURN MARK WHERE THE FIBER OPTIC CABLE HAD BEEN RESTING UPON IT. THERE WAS NO INJURY TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 592443 | SINGLE FIBER OPTIC CABLE, 10FT, PINK | SINGLE FIBER OPTIC CABLE, 10FT, PINK | FST | INVUITY, INC | FC1SP | UNKNOWN | 00816728020202 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|