FDA Adverse Event Malfunction Summary report: N

TRUEMETRIX

MDR report key: 7754614 · Received August 6, 2018

Report

Report Number
1000113657-2018-00834
Event Type
Malfunction
Date Received
August 6, 2018
Date of Event
July 12, 2018
Report Date
August 6, 2018
Manufacturer
TRIVIDIA HEALTH, INC.
Product Code
NBW
UDI-DI
00021292007041
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTERNAL REPORT # (B)(4). RETURNED METER EVALUATED WITH NO DEFECT FOUND. TEST STRIPS NOT RETURNED FOR EVALUATION. CUSTOMER RETURNED WRONG TEST STRIPS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-20 - USER'S TEST STRIP HAD POOR STORAGE (KITCHEN). TEST STRIP (B)(4). NOTE: MANUFACTURER CONTACTED CUSTOMER ON 7/26/2018 IN A FOLLOW-UP CALL TO ENSURE THAT THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN; CUSTOMER STATED THAT THE REPLACEMENT PRODUCT IS WORKING TO THEIR SATISFACTION AND THEY HAVE NOT HAD ANY MEDICAL INTERVENTION OR SYMPTOMS SINCE THE LAST CALL.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 180, 162, AND 171 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 80-120 MG/DL. CUSTOMER STATED THAT THE RESULTS HAVE BEEN HIGH SINCE OPENING A NEW BOTTLE OF STRIPS. CUSTOMER IS NOT USING ALCOHOL PADS OR SANITIZER. CUSTOMER HAS BEEN TESTING BOTH BEFORE A MEAL AND TWO HOURS AFTER. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2018, A BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULT OF 211 MG/DL USING TRUEMETRIX METER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 08/20/2019 AND OPEN VIAL DATE IS ONE WEEK. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
592918 TRUEMETRIX BLOOD GLUCOSE SYSTEM NBW TRIVIDIA HEALTH, INC. TRUEMETRIX MV2791 00021292007041

Patients

Seq Age Sex Outcome Treatment
1 0 YR