CER OPTION TYPE 1 TPR SLEVE -6
Report
- Report Number
- 0001825034-2018-05817
- Event Type
- Injury
- Date Received
- August 6, 2018
- Date of Event
- June 26, 2012
- Report Date
- October 11, 2018
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- LZO
- PMA / PMN Number
- PK082996
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. REPORTED EVENT WAS CONFIRMED BY REVIEW OF REVISION OPERATIVE NOTES PROVIDED. REVIEW OF OPERATIVE CONFIRMS THAT THE PATIENT UNDERWENT INCISION, DRAINAGE, DEBRIDEMENT, AND ONE STAGE REVISION OF RIGHT HIP ARTHROPLASTY REVISION DUE TO INFECTION. DURING THE PROCEDURE, ALL COMPONENTS WERE REMOVED AND NEW IMPLANTS WERE IMPLANTED. IT WAS NOTED THAT THE BONE WAS VERY HARD AND WHILE ATTEMPTING TO SEAT THE SCREW, THE SCREW BROKE. IT WAS DECIDED TO LEAVE THE BROKEN PORTION IN SITU, AS THE CUP WAS IN GOOD POSITION AND EXCELLENT POSITION FIXATION WITH THE TWO 20 SCREWS HAD BEEN OBTAINED. DEVICE HISTORY RECORD WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
(B)(4). CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION, AS THE DEVICE LOCATION IS UNKNOWN. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. UNIQUE IDENTIFIER (UDI) #: N/A. ITEM # 51-107120, STEM, LOT # 2483494. ITEM # 650-1057, HEAD, LOT # 423650. ITEM # 11-104062, SHELL, LOT # 160680. ITEM # EP-105996, LINER, LOT # 690770. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034 - 2018 - 05813. 0001825034 - 2018 - 05814. 0001825034 - 2018 - 05815. 0001825034 - 2018 - 05816.
IT WAS REPORTED THAT APPROXIMATELY 6 WEEKS POST IMPLANTATION, THE PATIENT UNDERWENT AN INCISION, DRAINAGE, DEBRIDEMENT, AND REMOVAL OF THE PROSTHESIS WITH A SINGLE STAGE EXCHANGE DUE TO INFECTION. THE PATIENT HAD BEEN EXPERIENCING DRAINAGE FROM THE WOUND OVER SEVERAL WEEKS AND ATTEMPTS TO TREAT WITH ORAL ANTIBIOTICS CONSERVATIVELY WERE UNSUCCESSFUL. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 593968 | CER OPTION TYPE 1 TPR SLEVE -6 | PROSTHESIS, HIP | LZO | ZIMMER BIOMET, INC. | N/A | 408520 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Hospitalization| R |