SUPERSHEATH
Report
- Report Number
- 9612126-2018-00010
- Event Type
- Injury
- Date Received
- August 6, 2018
- Date of Event
- July 6, 2018
- Manufacturer
- TOGO MEDIKIT CO. LTD.
- Product Code
- DRE
- PMA / PMN Number
- K141070
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE INVOLVED IN THE EVENT WAS RETURNED. THE RETURNED DEVICE WAS EXAMINED, AND THE REPEATABILITY TEST (TENSILE STRENGTH TESTING ETC.) WAS CONDUCTED USING RESERVED SAMPLES WITH THE LOT NUMBERS LISTED BELOW THAT HAD POSSIBILITIES TO BE THE SAME LOT NUMBER AS THAT INVOLVED IN THE EVENT. ALSO, THE INVESTIGATION WAS CONDUCTED BY REVIEWING THE RECORDS OF THE MANUFACTURING PROCESSES OF THE SHEATH WITH THE LOT NUMBERS LISTED BELOW, AND IT WAS CONFIRMED THAT THERE WERE NO MANUFACTURING PROCESSES THAT CAUSED OR CONTRIBUTED TO THE EVENT, AND THERE WERE NO MANUFACTURING RECORDS OF VISUAL INSPECTIONS THAT SHOWED THE CAUSE OF OR CONTRIBUTION TO THE EVENT. JUDGING FROM THIS EXAMINATION, A POSSIBLE CAUSE OF THIS FRACTURE IS THAT A SHARP EDGE SUCH AS A KNIFE CAME INTO CONTACT WITH THE SHEATH DURING THE PROCEDURE. THIS RESULTED IN A DECREASE IN TENSILE STRENGTH OF THE SHEATH TO A POINT WHERE THE SHEATH COULD NOT WITHSTAND THE PULL FORCE AND FRACTURED. THE MEDICAL DEVICE INVOLVED IN THE REPORTED EVENT IS SOLD ONLY IN (B)(4). THEREFORE, THE INFORMATION ON THE EQUIVALENT DEVICE TO THE COMPLAINT DEVICE THAT WAS CLEARED UNDER K141070. SUSPECT MEDICAL DEVICE" IN THIS REPORT. LOT#:18D13C8 AND 18C12C9. [(B)(4)].
ON (B)(6) 2018, AT A HOSPITAL IN (B)(6), IT WAS REPORTED THAT THE SUPER SHEATH WAS FRACTURED WHEN IT WAS PULLED OUT OF A PATIENT'S BODY DURING A PROCEDURE. THE FRACTURED PORTION REMAINED IN THE PATIENT'S BODY AND WAS REMOVED ADEQUATELY. THERE WAS NO REPORTED PATIENT INJURY AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 594466 | SUPERSHEATH | DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION | DRE | TOGO MEDIKIT CO. LTD. | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |