DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2018-086495
- Event Type
- Malfunction
- Date Received
- August 3, 2018
- Date of Event
- May 22, 2018
- Report Date
- October 15, 2018
- Manufacturer
- DEXCOM, INC.
- Product Code
- PQF
- UDI-DI
- 10386270000221
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4)
DEXCOM WAS MADE AWARE ON 07/07/2018 THAT THE DISPLAY DEVICE SHOWED A TRANSMITTER FAILED ERROR ON (B)(6) 2018. THE DEVICE HAS BEEN RECEIVED FOR INVESTIGATION. A FOLLOW-UP WILL BE SUBMITTED UPON COMPLETION OF DEVICE INVESTIGATION. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.
DEXCOM WAS MADE AWARE ON 07/07/2018 THAT THE DISPLAY DEVICE SHOWED A TRANSMITTER FAILED ERROR ON (B)(6) 2018. THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. VOLTAGE TESTING WAS PERFORMED, AND THE TEST FAILED. A REVIEW OF THE DOWNLOADED DATA LOG CONFIRMED THE REPORTED EVENT OF A FAILED TRANSMITTER ERROR. THE PROBABLE CAUSE WAS DETERMINED TO BE A LOW TRANSMITTER BATTERY. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 588089 | DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | PQF | DEXCOM, INC. | 9438-06 | 6007249 | 10386270000221 |