FDA Adverse Event Injury Summary report: N

LCP(TM) DISTAL FEMUR PLATE 11 HOLES/276MM-RIGHT

MDR report key: 7742405 · Received August 2, 2018

Report

Report Number
8030965-2018-55439
Event Type
Injury
Date Received
August 2, 2018
Report Date
July 12, 2018
Manufacturer
OBERDORF SYNTHES PRODUKTIONS GMBH
Product Code
HRS
UDI-DI
07611819189126
PMA / PMN Number
K062564
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. ADDITIONAL NARRATIVE: METHOD CODES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. DHR REVIEW WAS COMPLETED. PART: 222.256 . LOT: 9796925 . MANUFACTURING SITE: MEZZOVICCO . RELEASE TO WAREHOUSE DATE: 15 JAN.2016 . THE DEVICE HISTORY RECORD SHOWS THIS LOT OF 12 PIECES WAS PROCESSED THROUGH THE NORMAL MANUFACTURING AND INSPECTION OPERATIONS WITH NO REWORK OR NONCONFORMITIES NOTED. THIS LOT MET ALL DIMENSIONAL AND VISUAL CRITERIA AT THE TIME OF RELEASE WITH NO ISSUES DOCUMENTED DURING THE MANUFACTURING PROCESS. THE RAW MATERIAL CERTIFICATE 18944 WAS REVIEWED AND THE USED MATERIAL WAS ACCORDING TO ISO-5832-1 SPECIFICATION FOR IMPLANTS FOR SURGERY REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION CUSTOMER QUALITY CONDUCTED AN INVESTIGATION OF THE RETURNED DEVICE. A POST-OPERATIVELY BROKEN LCP-DF 4.5/5 R 11HO L276 SST WAS RETURNED FOR INVESTIGATION. THE IMPLANT WAS FORWARDED TO THE MANUFACTURING SITE (SYNTHES MEZZOVICCO) AND WAS CHECKED FOR CONFORMANCE TO THE SPECIFICATIONS. THE REVIEW OF THE DEVICE HISTORY RECORD REVEALED THAT THIS LCP-DF 4.5/5 R 11HO L276 SST WAS MANUFACTURED IN JANUARY 2016 ACCORDING TO THE SPECIFICATIONS. THE PARTS CONFORMED TO DIMENSIONAL SPECIFICATIONS AT THE TIME OF MANUFACTURING AND PASSED INSPECTION REQUIREMENTS WITH NO NON-CONFORMITIES REPORTED. A VISUAL INSPECTION HAS BEEN PERFORMED AND IT CONFIRMED THAT THE RECEIVED CONDITION AGREES WITH THE COMPLAINT CONDITION BUT NO VISUAL DEFECTS MANUFACTURING RELATED HAVE BEEN IDENTIFIED ON RETURNED ITEM. THE DIMENSIONAL RE-INSPECTION, THE RAW MATERIAL CERTIFICATE REVIEW AND THE DOCUMENT/SPECIFICATION REVIEW DIDN¿T IDENTIFY ANY MANUFACTURING DEFECT OR DEFICIENCY, THUS THE COMPLAINT INVESTIGATION IS CONSIDERED AS NOT MANUFACTURING RELATED. THE APPLICABLE RISK MANAGEMENT DOCUMENT WAS REVIEWED AND FOUND TO ADEQUATELY ADDRESS THIS COMPLAINT CONDITION. NO MANUFACTURING RELATED ISSUE WAS IDENTIFIED, THEREFORE REVIEW TO THE SPECIFIC PRM AND PRM LINE IS NOT APPLICABLE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

PATIENT DATE OF BIRTH AND WEIGHT NOT AVAILABLE FOR REPORTING. DATE OF DEVICE BREAKAGE IS NOT KNOWN. INITIAL REPORTER TELEPHONE NOT AVAILABLE. INITIAL REPORTER IS SYNTHES AFFILIATE. A REVIEW OF THE DEVICE HISTORY RECORDS HAS BEEN REQUESTED. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED PATIENT WAS IMPLANTED WITH A LOCKING COMPRESSION PLATE (LCP) DISTAL FEMUR PLATE AND SCREW ON (B)(6) 2018. X-RAY TAKEN ON (B)(6) 2018 REVEALED THE PLATE WAS BROKEN. PATIENT WAS RETURNED TO SURGERY ON THE SAME DATE FOR REVISION SURGERY. PATIENT STATUS REPORTED AS STABLE. CONCOMITANT DEVICES REPORTED: 5.0MM LOCKING SCREW SELF-TAPPING 34MM (213.334, LOT L625766, QUANTITY 1), 5.0MM LOCKING SCREW SELF-TAPPING 36MM (213.336, LOT 9763925, QUANTITY 1), 5.0MM LOCKING SCREW SELF-TAPPING 36MM (213.336, LOT L277973, QUANTITY 1), 5.0MM LOCKING SCREW SELF-TAPPING 38MM (213.338, LOT L241635, QUANTITY 1), 5.0MM LOCKING SCREW SELF-TAPPING 46MM (213.346, LOT 8593682, QUANTITY 1), 5.0MM LOCKING SCREW SELF-TAPPING 46MM (213.346, LOT 9754055, QUANTITY 1), 5.0MM LOCKING SCREW SELF-TAPPING 55MM (213.355, LOT L570704, QUANTITY 1), 5.0MM LOCKING SCREW SELF-TAPPING 65MM (213.365, LOT L631292, QUANTITY 2), 5.0MM LOCKING SCREW SELF-TAPPING 75MM (213.375, LOT L471084, QUANTITY 1), 5.0MM LOCKING SCREW SELF-TAPPING 75MM (213.375, LOT L487983, QUANTITY 1), CABLE/WIRE (PART NUMBER UNKNOWN, LOT NUMBER UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) LCP DISTAL FEMUR PLATE. THIS IS REPORT 1 OF 2 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
585052 LCP(TM) DISTAL FEMUR PLATE 11 HOLES/276MM-RIGHT PLATE, FIXATION, BONE HRS OBERDORF SYNTHES PRODUKTIONS GMBH 9796925 07611819189126

Patients

Seq Age Sex Outcome Treatment
1 85 YR Required Intervention