SOFTWARE FOR DIAGNOSIS/TREATMENT
Report
- Report Number
- 1061932-2018-01036
- Event Type
- Malfunction
- Date Received
- August 1, 2018
- Date of Event
- July 2, 2018
- Report Date
- October 15, 2018
- Manufacturer
- BECKMAN COULTER
- Product Code
- OSN
- Removal / Correction Number
- 2050012-0911/2018-013C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
FOLLOW UP 01: FIELD ACTION (FA-000072-2) WAS INITIATED TO ADDRESS THIS ISSUE. INSTALLATION OF KALUZA VERSION 2.1.1 WILL PREVENT SOFTWARE CRASHES AND RESULT IN PROPER KEYWORD VALUE BEING TRANSMITTED FOR ALL SAMPLES. KALUZA ANALYSIS 2.1.1 CAN BE DOWNLOADED FROM THE BECKMAN COULTER WEBSITE AND NEEDS TO BE INSTALLED ON ALL AFFECTED SYSTEMS. ADDITIONAL INFORMATION HAS BEEN SELECTED. NOTIFICATION AND MODIFICATION HAVE BEEN SELECTED. NUMBER HAS BEEN UPDATED TO REFLECT THE RESPECTIVE INFORMATION FOR THIS EVENT. ADDITIONAL INFORMATION: FA-000072-2 INCLUDES NOTIFICATION TO THE CUSTOMER AND CORRECTION OF THE ISSUE VIA SOFTWARE UPDATE. BEC INTERNAL IDENTIFIER - (B)(4).
KALUZA 2.0 SOFTWARE IS AN RUO (RESEARCH USE ONLY) PRODUCT; THEREFORE THERE IS NO IMPACT TO PATIENT RESULTS FOR THIS EVENT. HOWEVER, THERE IS A SIMILAR PRODUCT AVAILABLE WHICH IS LABELED FOR CLINICAL USE (IVD), KALUZA C, 10 USER NETWORK EDUCATIONAL LICENSE SOFTWARE KIT (PN: C10580). THE TECHNICAL SUPPORT SPECIALIST CONFIRMED THE PROBLEM FILED THIS COMPLAINT AND ESCALATED THE ISSUE TO THE SOFTWARE ENGINEER. THE PROBLEM WAS CONFIRMED BY THE SOFTWARE ENGINEERING TEAM AND EXTENDS TO ALL RUO LICENSE TYPES AND CLINICAL VERSIONS OF THE STAND-ALONE ANALYSIS SOFTWARE. THE FOLLOWING WAS OBSERVED: WHEN A NEW FILE IS UPLOADED THE KEYWORDS CREATED IN THE CYTOMETER SOFTWARE ARE CORRECTLY UPDATED IN THE INFO PLOTS. THE KEYWORD VALUES ON THE LIS PREVIEW/EXPORT SCREEN NEVER UPDATE AFTER THE TEMPLATE IS FIRST CREATED. THE FIRST VALUES ENTERED ARE ALWAYS DISPLAYED. SECONDARY SAMPLE IDENTIFIERS (IDS) CAN BE INCLUDED AS KEYWORDS AND WOULD BE TRANSMITTED INCORRECTLY; HOWEVER THE PRIMARY SAMPLE ID IS NOT CODED AS A KEYWORD AND WOULD NOT BE IMPACTED. A STOP SHIP (MIA-18-0257-Q) WAS IMPLEMENTED ON 20-JULY-2018 FOR BOTH RUO AND CLINICAL VERSIONS OF THE SOFTWARE. THE ISSUE WAS ESCALATED TO POST MARKET FOR ASSESSMENT.
THE CUSTOMER REPORTED A PROBLEM WITH KALUZA 2.0 SOFTWARE IN WHICH KEYWORDS ENTERED BY THE CUSTOMER DID NOT PROPERLY TRANSFER TO THE LABORATORY INFORMATION SYSTEM (LIS). THE LIS TEMPLATE SAVES THE KEYWORD INFORMATION OF THE FILE LOADED WHEN THE TEMPLATE WAS CREATED. WHEN A NEW FILE IS DRAGGED IN, THE KEYWORDS ARE NOT UPDATED. CUSTOMERS ARE PLACING INFORMATION NEEDED FOR RESULT CALCULATION AND FILING IN THE SAMPLE ID FIELDS FOR TRANSMISSION TO LIS. THESE FIELDS (SUCH AS CBC, LYMPH %, DILUTION FACTOR, SAMPLE ID, SERIAL #) ARE THEN USED ON LIS TO CALCULATE FINAL PATIENT RESULT. IF A CUSTOMER DOES NOT REALIZE THE FIELDS TRANSMITTED ARE FOR WRONG DATA FILE, THE RESULTS WILL BE INCORRECT. KALUZA 2.0 IS AN RUO (RESEARCH USE ONLY) PRODUCT; THEREFORE THERE IS NO IMPACT TO PATIENT RESULTS FOR THIS EVENT. HOWEVER, THERE IS A SIMILAR PRODUCT AVAILABLE WHICH IS LABELED FOR CLINICAL USE (IVD), KALUZA C, 10 USER NETWORK EDUCATIONAL LICENSE SOFTWARE KIT (PN: C10580).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 580108 | SOFTWARE FOR DIAGNOSIS/TREATMENT | OSN | BECKMAN COULTER | SWRE KIT,KALUZA C 10 USER NETWORK EDUCATIONAL LICENSE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |