HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY
Report
- Report Number
- 3007042319-2018-03408
- Event Type
- Malfunction
- Date Received
- August 1, 2018
- Date of Event
- November 21, 2017
- Report Date
- October 4, 2018
- Manufacturer
- HEARTWARE, INC.
- Product Code
- DSQ
- UDI-DI
- 00888707000369
- PMA / PMN Number
- P100047
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
PRODUCT EVENT SUMMARY: THE BATTERY WAS RETURNED FOR EVALUATION. VARIOUS ANALYSES WERE CONDUCTED AND REVIEWED IN ORDER TO EVALUATE THE PERFORMANCE OF THE DEVICES IN RELATION TO THE REPORTED EVENT. FAILURE ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE DEVICE PASSED VISUAL EXAMINATION AND FUNCTIONAL TESTING. A REVIEW OF THE BATTERY'S INTERNAL LOG REVEALED THAT A CELL PAIR, AT ONE POINT IN TIME, PASSED THE VOLTAGE THRESHOLD. WHEN THIS OCCURS, THE BATTERY WILL NOT CHARGE UNTIL THE VOLTAGE DECREASES BELOW THE THRESHOLD. HOWEVER, THE BATTERY WAS RECEIVED WITH NO FLAGS ENABLED. IF THE BATTERY REMAINS CONNECTED TO THE BATTERY CHARGER WITH A CELL-OVER-VOLTAGE FLAG, THE BATTERY CHARGER STATUS INDICATOR WILL FLASH RED AFTER 8 HOURS DUE TO A CHARGE TIME-OUT. AS A RESULT, THE REPORTED EVENT WAS CONFIRMED. THE MOST LIKELY ROOT CAUSE OF THE REPORTED BATTERY NOT CHARGING CAN BE ATTRIBUTED TO AN OVERVOLTAGE FAULT BY THE ANALOG FRONT END (AFE) INTEGRATED CIRCUIT. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
PRODUCT EVENT SUMMARY: THE RETURNED BATTERY PASSED VISUAL INSPECTION AND FUNCTIONAL TESTING. THE REPORTED EVENT COULD NOT BE CONFIRMED DURING BENCH TESTING. INVESTIGATION IS ONGOING. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE BATTERY WAS UNABLE TO BE CHARGED AND COULD NOT PROVIDE POWER. THE BATTERY WAS REPLACED. THE BATTERY WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 580107 | HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY | VENTRICULAR (ASSISST) BYPASS | DSQ | HEARTWARE, INC. | 1650DE | 00888707000369 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | 1103 VAD |