GALILEO
Report
- Report Number
- 1034569-2006-00084
- Event Type
- Malfunction
- Date Received
- August 23, 2006
- Date of Event
- July 25, 2006
- Report Date
- August 23, 2006
- Manufacturer
- IMMUCOR
- Product Code
- KSZ
- PMA / PMN Number
- bk040013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
ONLY THE CUSTOMER SAMPLE REPORTED TO HAVE ANTI-JKA WAS SENT FOR INVESTIGATION. THE PRESENCE OF THE E AND JKA ANTIGENS WERE CONFIRMED ON CAPTURE-R READY-ID, LOT ID073. TESTING WAS PERFORMED WITH THE CUSTOMER'S SAMPLE USING RETENTION CAPTURE-R REASY-ID, LOT ID073 AND CAPTURE-R INDICATOR RED CELLS, LOT 221897; THE SAMPLE WAS NONREACTIVE WITH ALL CELLS TESTED. IN-HOUSE DONOR SAMPLES AND THE CUSTOMER'S SUBMITTED SAMPLE WERE TESTED ON AN IN-HOUSE GALILEO. THE IN-HOUSE DONOR SAMPLES WERE NONREACTIVE, AS EXPECTED. THE CUSTOMER'S SAMPLE WAS REACTIVE. HEMAGGLUTINATION TUBE TESTING WAS PERFORMED WITH CUSTOMER'S SAMPLE USING RETENTION PANOSCREEN, LOT 23248. THE SAMPLE EXHIBITED 1+ MACROSCOPIC REACTIVITY WITH JK (A+B-) CELLS, WEAK MACROSCOPIC REACTIVITY WITH JK (A+B+) CELLS AND WAS NONREACTIVE WITH JK (A-) CELLS. THE GALILEO OPERATORS MANUAL STATES, "THE GALILEO CANNOT RELIABLY DETECT HEMAGGLUTINATION REACTIONS THAT ARE GRADED AS 1+ OR LESS IN TEST TUBE METHODOLOGY." THE CUSTOMER'S COMPLAINT APPEAR TO BE RELATED TO THE NATURE OF THE SPECIMEN USED FOR TESTING. ALTHOUGH A TRANSFUSION OF INCOMPATIBLE BLOOD DID NOT OCCUR, THE POTENTIAL FOR TRANSFUSION OF INCOMPATIBLE BLOOD EXISTS.
CUSTOMER REPORTED MISSED ANTIBODIES ON THE GALILEO INSTRUMENT. TWO SAMPLES FROM DIFFERENT PTS HAD ANTI-JKA AND ANTI-E, RESPECTIVELY, WHICH WERE NOT DETECTED BY THE GALILEO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GALILEO | KSZ | KSZ | IMMUCOR | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | CAPTURE-R INDICATOR RED CELLS, LOT 221897| CAPTURE-R READY-ID, LOT ID073 |