FDA Adverse Event Malfunction Summary report: N

GALILEO

MDR report key: 774071 · Received August 23, 2006

Report

Report Number
1034569-2006-00084
Event Type
Malfunction
Date Received
August 23, 2006
Date of Event
July 25, 2006
Report Date
August 23, 2006
Manufacturer
IMMUCOR
Product Code
KSZ
PMA / PMN Number
bk040013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ONLY THE CUSTOMER SAMPLE REPORTED TO HAVE ANTI-JKA WAS SENT FOR INVESTIGATION. THE PRESENCE OF THE E AND JKA ANTIGENS WERE CONFIRMED ON CAPTURE-R READY-ID, LOT ID073. TESTING WAS PERFORMED WITH THE CUSTOMER'S SAMPLE USING RETENTION CAPTURE-R REASY-ID, LOT ID073 AND CAPTURE-R INDICATOR RED CELLS, LOT 221897; THE SAMPLE WAS NONREACTIVE WITH ALL CELLS TESTED. IN-HOUSE DONOR SAMPLES AND THE CUSTOMER'S SUBMITTED SAMPLE WERE TESTED ON AN IN-HOUSE GALILEO. THE IN-HOUSE DONOR SAMPLES WERE NONREACTIVE, AS EXPECTED. THE CUSTOMER'S SAMPLE WAS REACTIVE. HEMAGGLUTINATION TUBE TESTING WAS PERFORMED WITH CUSTOMER'S SAMPLE USING RETENTION PANOSCREEN, LOT 23248. THE SAMPLE EXHIBITED 1+ MACROSCOPIC REACTIVITY WITH JK (A+B-) CELLS, WEAK MACROSCOPIC REACTIVITY WITH JK (A+B+) CELLS AND WAS NONREACTIVE WITH JK (A-) CELLS. THE GALILEO OPERATORS MANUAL STATES, "THE GALILEO CANNOT RELIABLY DETECT HEMAGGLUTINATION REACTIONS THAT ARE GRADED AS 1+ OR LESS IN TEST TUBE METHODOLOGY." THE CUSTOMER'S COMPLAINT APPEAR TO BE RELATED TO THE NATURE OF THE SPECIMEN USED FOR TESTING. ALTHOUGH A TRANSFUSION OF INCOMPATIBLE BLOOD DID NOT OCCUR, THE POTENTIAL FOR TRANSFUSION OF INCOMPATIBLE BLOOD EXISTS.

Description of Event or Problem · 1

CUSTOMER REPORTED MISSED ANTIBODIES ON THE GALILEO INSTRUMENT. TWO SAMPLES FROM DIFFERENT PTS HAD ANTI-JKA AND ANTI-E, RESPECTIVELY, WHICH WERE NOT DETECTED BY THE GALILEO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GALILEO KSZ KSZ IMMUCOR * *

Patients

Seq Age Sex Outcome Treatment
1 * CAPTURE-R INDICATOR RED CELLS, LOT 221897| CAPTURE-R READY-ID, LOT ID073