MENTOR SMOOTH ROUND MODERATE PROFILE
Report
- Report Number
- 1645337-2018-04634
- Event Type
- Injury
- Date Received
- July 31, 2018
- Date of Event
- July 2, 2018
- Report Date
- July 9, 2018
- Manufacturer
- MENTOR TEXAS
- Product Code
- FWM
- UDI-DI
- 00081317001294
- PMA / PMN Number
- P990075
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITHOUT FLUID. YELLOW MATERIAL WAS OBSERVED WITHIN THE DEVICE. NO FOREIGN MATERIAL WAS OBSERVED ON THE SHELL SURFACE. PRODUCT EVALUATION DISCOVERED A RENT MEASURING APPROXIMATELY 0.2 CM WITHIN A CREASE ON THE ANTERIOR ASPECT. ALSO WAS NOTICED SEVERAL CREASES ON BOTH VIEWS ASPECTS OF THE DEVICE. NO OTHER ANOMALIES WERE DISCOVERED. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE. THE DEVICE HISTORY RECORD ( DHR) OF LOT NUMBER 5951688 WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. CONCOMITANT PRODUCTS: MENTOR SMOOTH ROUND MODERATE PROFILE 475 CC SALINE, CATALOG: 3501680, LOT: 5940912 SN: (B)(4). (B)(4).
IT WAS REPORTED THAT A (B)(6)-YEAR OLD CAUCASIAN FEMALE PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH MENTOR SMOOTH ROUND MODERATE PROFILE 475 CC SALINE BREAST PROSTHESES. DEFLATION OF THE RIGHT BREAST PROSTHESIS WAS IDENTIFIED. AS A RESULT, THE PATIENT UNDERWENT BILATERAL REMOVAL AND REPLACEMENT WITH MENTOR MEMORY GEL BREAST IMPLANT 500 CC GEL BREAST PROSTHESES ON (B)(6) 2018.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 578299 | MENTOR SMOOTH ROUND MODERATE PROFILE | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | MENTOR TEXAS | 5951688 | 00081317001294 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR | Required Intervention |