FDA Adverse Event Malfunction Summary report: N

HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0

MDR report key: 7734622 · Received July 31, 2018

Report

Report Number
3007042319-2018-03353
Event Type
Malfunction
Date Received
July 31, 2018
Date of Event
June 22, 2018
Report Date
August 22, 2019
Manufacturer
HEARTWARE, INC.
Product Code
DSQ
UDI-DI
00888707000420
PMA / PMN Number
P100047
Removal / Correction Number
Z-1903-2018
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

PRODUCT EVENT SUMMARY: ONE CONTROLLER, SIX BATTERIES AND TWO CONTROLLER AC ADAPTERS WERE NOT RETURNED FOR EVALUATION. LOG FILE ANALYSIS REVEALED THAT THE CONTROLLER, (B)(4), CONTAINED A FEATURE THAT RECORDS WHETHER A POWER SOURCE EXPERIENCED A COMMUNICATION ERROR OR A DISCONNECTION WITHIN EACH 15-MINUTE INTERVAL. ANALYSIS OF THE DATA LOG FILES REVEALED SEVERAL PREMATURE POWER SWITCHING EVENTS THAT WERE DUE TO MOMENTARY DISCONNECTIONS INVOLVING (B)(4) AND PREMATURE POWER SWITCHING EVENTS THAT WERE DUE TO COMMUNICATION ERRORS INVOLVING (B)(4). AS A RESULT, THE REPORTED EVENT WAS CONFIRMED. A POWER SOURCE LUBRICATION PROCEDURE WAS PERFORMED ON THE BATTERIES AND CONTROLLER AC ADAPTERS IN ACCORDANCE WITH THE REQUIREMENTS UNDER FSCA (B)(4) ON JULY 6, 2018 TO MITIGATE THE REPORTED CONDITIONS AND THE DEVICES REMAIN IN USE. THE MOST LIKELY ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS AND COMMUNICATION ERRORS BETWEEN THE CONTROLLER AND BATTERIES. AN INTERNAL INVESTIGATION HAS BEEN OPENED TO EVALUATE MOMENTARY DISCONNECTIONS AND IMPLEMENT CORRECTIVE ACTIONS AS REQUIRED. CORRECTION FOR PREVIOUS REPORT: CAC016144 ¿ CATALOGUE AND MODEL NUMBER ¿ 1430DE (B)(4). MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

PRODUCT EVENT SUMMARY: THE CONTROLLER, SIX BATTERIES, AND TWO CONTROLLER AC ADAPTERS WERE NOT RETURNED FOR EVALUATION. LOG FILE ANALYSIS REVEALED THAT THE CONTROLLER CONTAINED A FEATURE THAT RECORDS WHETHER A POWER SOURCE EXPERIENCED A COMMUNICATION ERROR OR A DISCONNECTION WITHIN EACH 15-MINUTE INTERVAL. DATA LOG FILES REVEALED MULTIPLE INSTANCES INVOLVING BAT587157, BAT587158, BAT589249, BAT587162 AND BAT587163 WHERE THE BATTERIES¿ RELATIVE STATE OF CHARGE (RSOC) WAS BETWEEN 101-201, WHICH ARE INDICATIVE OF COMMUNICATION ERRORS. ANALYSIS OF THE ALARM LOG FILE REVEALED ONE (1) CRITICAL BATTERY ALARM DUE TO COMMUNICATION ERRORS INVOLVING BAT587157. ANALYSIS OF THE DATA LOG FILES REVEALED SEVERAL PREMATURE POWER SWITCHING EVENTS THAT WERE DUE TO MOMENTARY DISCONNECTIONS INVOLVING BAT589249, BAT587162, BAT587157, BAT587156 AND PREMATURE POWER SWITCHING EVENTS THAT WERE DUE TO COMMUNICATION ERRORS INVOLVING BAT587162 AND BAT587157. AS A RESULT, THE REPORTED EVENT WAS CONFIRMED. A POWER SOURCE LUBRICATION PROCEDURE WAS PERFORMED ON THE BATTERIES AND CONTROLLER AC ADAPTERS ON 6/JUL/2018 TO MITIGATE THE REPORTED CONDITIONS, AND THE DEVICES REMAIN IN USE. THE MOST LIKELY ROOT CAUSE OF THE PREMATURE POWER SWITCHING EVENTS CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS AND COMMUNICATION ERRORS BETWEEN THE CONTROLLER AND BATTERIES. THE MOST LIKELY ROOT CAUSE OF THE CRITICAL BATTERY ALARMS CAN BE ATTRIBUTED TO COMMUNICATION ERRORS BETWEEN THE CONTROLLER AND BATTERIES. THE MOST LIKELY ROOT CAUSE OF THE REPORTED RSOC VALUES BETWEEN 101-201 CAN BE ATTRIBUTED TO COMMUNICATION ERRORS BETWEEN THE CONTROLLER AND BATTERIES. POSSIBLE ROOT CAUSES OF THE COMMUNICATION ERRORS CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS ON THE COMMUNICATION PINS OF THE CONTROLLER, THE CONTROLLER NOT RECEIVING RESPONSES FROM THE BATTERY, AND/OR DUE TO THE PACKET ERROR CHECKING METHOD DETECTING BIT ERRORS. THIS EVENT WAS ASSESSED AND IS BEING REPORTED AS PART OF A RETROSPECTIVE REVIEW OF LOG FILE DATA. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

CORRECTED FIELD: THIS EVENT WAS ASSESSED AND IS BEING REPORTED AS PART OF A RETROSPECTIVE REVIEW OF LOG FILE DATA. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

THE SITE FURTHER REPORTED THAT A MANUFACTURER REPRESENTATIVE SERVICED THE ALLEGED DEVICES AND THE DEVICES ARE STILL IN USE.

Description of Event or Problem · 0

LOG FILE REVIEW INDICATED THAT FOUR BATTERIES EXHIBITED COMMUNICATION ERRORS.

Description of Event or Problem · 0

LOG FILE REVIEW INDICATED THAT ONE BATTERY EXHIBITED A CRITICAL BATTERY ALARM.

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION AND DESTINATION THERAPY IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. OTHER DEVICES INVOLVED IN THIS EVENT: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4) / MODEL #: 1650DE / EXPIRATION DATE: 2018-10-31 UDI #: (B)(4) MFG DATE: 2017-10-31 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2018-10-31 UDI #: (B)(4) MFG DATE: 2017-10-31 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2018-11-30 UDI #:(B)(4) MFG DATE: 2017-11-20 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY/ (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2018-10-31 UDI #: (B)(4) MFG DATE: 2017-10-31 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2018-10-31 UDI #:(B)(4) MFG DATE: 2017-10-31 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4) / MODEL #: 1650DE / EXPIRATION DATE: 2018-10-31 UDI #: (B)(4) MFG DATE: 2017-10-31 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ CAC ADAPER / (B)(4)/ MODEL #: UNK-CAC ADAPTER (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ CAC ADAPTER / (B)(4)/ MODEL #: 1430DE UDI #: (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT THE PATIENT EXPERIENCED POWER SWITCHING WHILE AT HOME. THE SITE FURTHER REPORTED THAT THE CHARGE ON THE BATTERIES AT THE TIME OF THE EVENT WAS GREATER THAN OR EQUAL TO TWENTY-FIVE PERCENT. THE DEVICES ARE STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
578935 HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0 VENTRICULAR (ASSIST) BYPASS DSQ HEARTWARE, INC. 1420 00888707000420

Patients

Seq Age Sex Outcome Treatment
1 48 YR MCS UNKNOWN VAD.