FDA Adverse Event Malfunction Summary report: N

SITTER ELITE

MDR report key: 7733526 · Received July 30, 2018

Report

Report Number
2020362-2018-00061
Event Type
Malfunction
Date Received
July 30, 2018
Report Date
September 13, 2018
Manufacturer
POSEY PRODUCTS LLC
Product Code
PJO
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RECEIVED IN WITH SCRATCHES TO THE EXTERIOR HOUSING. DAMAGES TO THE HOUSING AND DUST COVERS IN THE BATTERY SLOT WAS ALSO OBSERVED. FUNCTIONAL TESTING SHOWED THE DEVICE DID NOT SOUND WHEN IN USE WITH THE PULL MAGNET AND SENSOR PAD. UPON OPENING THE DEVICE, THE CAUSE FOR THE FAILURE WAS A FAULTY SPEAKER, WHERE THE SPEAKER IMPEDANCE WAS OUT OF RANGE. THIS LOSS OF FUNCTIONALITY WOULD RESULT IN THE ALARM¿S FAILURE TO ALERT THE CAREGIVER OF A PATIENT EXIT AND COULD CONTRIBUTE TO A PATIENT INCIDENT. BASE ON THE AGE OF THE DEVICE, IT IS LIKELY EXPECTED NORMAL WEAR AND TEAR CONTRIBUTED TO THE DEFECT. AT THIS TIME, THERE IS NO EVIDENCE THAT A MANUFACTURING NON-CONFORMITY CONTRIBUTED TO THE REPORTED COMPLAINT AND THE INSTRUCTIONS FOR USE (IFU) WERE REVIEWED AND DETERMINED TO PROVIDE ADEQUATE INSTRUCTIONS AND WARNING FOR THE SAFE AND EFFECTIVE USE OF THE DEVICE. THE IFU STATES, TO REDUCE THE RISK OF SERIOUS INJURY OR DEATH, TEST THE ALARM AND SENSOR FOR PROPER OPERATION PRIOR TO PUTTING IN SERVICE WITH A PATIENT, AND EACH TIME BEFORE LEAVING THE PATIENT UNATTENDED. IF THE ALARM AND/OR SENSOR DO NOT FUNCTION PROPERLY, REMOVE THE ALARM AND SENSOR FROM SERVICE AND REPLACE THEM WITH A PROPERLY FUNCTIONING ALARM AND/OR SENSOR. DO NOT USE THE ALARM, SENSOR OR MAGNET IF IT DOES NOT ACTIVATE EACH TIME WEIGHT IS REMOVED FROM THE SENSOR, THE CHAIR BELT SENSOR IS UNFASTENED, OR MAGNET IS REMOVED FROM FACE PLATE. THEREFORE, NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE NECESSARY. ALL COMPLAINTS ARE TRENDED AND REVIEWED BY MANAGEMENT ON A MONTHLY BASIS. AS PART OF THIS MONTHLY REVIEW, ANY EXCURSION ABOVE THE CONTROL LIMITS FOR THIS FAILURE MODE WILL BE ASSESSED, DOCUMENTED AND ACTED UPON AS WARRANT. MANUFACTURER REFERENCE FILE #(B)(4).

Additional Manufacturer Narrative · 1

PRODUCT IS SCHEDULED TO BE RETURNED BUT HAVE NOT BEEN RECEIVED IN BY MANUFACTURING AT THE TIME OF THIS REPORT. THEREFORE, THIS REPORT IS BASED SOLELY ON THE INFORMATION PROVIDED BY THE CUSTOMER. AT THIS TIME, THERE IS NO EVIDENCE THAT A MANUFACTURING NON-CONFORMITY CONTRIBUTED TO THE REPORTED COMPLAINT AND THAT INSTRUCTIONS FOR USE (IFU) WERE REVIEWED AND DETERMINED TO PROVIDE ADEQUATE INSTRUCTIONS AND WARNING FOR THE SAFE AND EFFECTIVE USE OF THE DEVICE. THE IFU STATES "TEST THE ALARM AND SENSOR FOR PROPER OPERATION PRIOR TO PUTTING IN SERVICE WITH A PATIENT, AND EACH TIME BEFORE LEAVING THE PATIENT UNATTENDED. IF THE ALARM AND/OR SENSOR DO NOT FUNCTION PROPERLY, REMOVE THE ALARM AND SENSOR FROM SERVICE AND REPLACE THEM WITH A PROPERLY FUNCTIONING ALARM AND/OR SENSOR. DO NOT USE THE ALARM, SENSOR OR MAGNET IF IT DOES NOT ACTIVATE EACH TIME WEIGHT IS REMOVED FROM THE SENSOR, THE CHAIR BELT SENSOR IS UNFASTENED, OR MAGNET IS REMOVED FROM FACE PLATE." THEREFORE, NO CORRECTIVE OR PREVENTIVE ACTIONS ARE NECESSARY. ALL COMPLAINTS ARE TRENDED AND REVIEWED BY MANAGEMENT ON A MONTHLY BASIS. AS PART OF THIS MONTHLY REVIEW, ANY EXCURSION ABOVE THE CONTROL LIMITS FOR THIS FAILURE MODE WILL BE ASSESSED, DOCUMENTED AND ACTED UPON AS WARRANTED. (B)(4). PENDING DEVICE RETURN.

Description of Event or Problem · 1

CUSTOMER REPORTED THE ALARM DOES NOT SOUND WHEN IT SHOULD. CUSTOMER REPORTED THE TONE LIGHT FLASHES. THE DATE THE ISSUE WAS DISCOVERED IS UNKNOWN AND NO PATIENT INCIDENT OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
574270 SITTER ELITE FALL PREVENTION ALARM/SENSOR COMBINATION ATTACHED OR UNATTACHED PJO POSEY PRODUCTS LLC 8345

Patients

Seq Age Sex Outcome Treatment
1