FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PROFILE 475CC

MDR report key: 7733309 · Received July 30, 2018

Report

Report Number
1645337-2018-04601
Event Type
Injury
Date Received
July 30, 2018
Date of Event
June 27, 2018
Report Date
July 6, 2018
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001294
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON 08/03/2018, MENTOR COMPLETED THE INVESTIGATION ON THE SUSPECT MEDICAL DEVICE. UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITHOUT FLUID. NO FOREIGN MATERIAL WAS OBSERVED WITHIN THE DEVICE OR ON THE SHELL SURFACE. THE PRODUCT EVALUATION (PE) TEAM DISCOVERED A RENT AT THE PATCH'S JUNCTURE. IN ADDITION, A CREASE WAS NOTICED ON THE ANTERIOR ASPECT. NO OTHER ANOMALIES WERE DISCOVERED. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET; HOWEVER, DEFLATION TRENDS FOR MENTOR SALINE-FILLED DEVICES WILL CONTINUE TO BE MONITORED BY QUALITY ASSURANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6)-YEAR-OLD FEMALE PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION WITH MENTOR SMOOTH ROUND MODERATE PROFILE 475CC SALINE BREAST IMPLANTS WHICH THE RIGHT SIDE DEFLATED AFTER IMPLANTATION. THE ISSUE WAS DIAGNOSED DURING THE PHYSICAL EXAMINATION. AS A RESULT, PATIENT HAD THE DEVICE REMOVED AND REPLACED WITH CATALOG NUMBER 3501680, SERIAL NUMBER (B)(4), LOT NUMBER 6853770 ON (B)(6) 2018.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
574075 MENTOR SMOOTH ROUND MODERATE PROFILE 475CC PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 5697122 00081317001294

Patients

Seq Age Sex Outcome Treatment
1 39 YR Required Intervention