VITEK® 2 AST-GN70 TEST KIT
Report
- Report Number
- 1950204-2018-00275
- Event Type
- Malfunction
- Date Received
- July 30, 2018
- Report Date
- September 16, 2018
- Manufacturer
- BIOMERIEUX, INC
- Product Code
- LON
- UDI-DI
- 03573026398477
- PMA / PMN Number
- K152075
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A BIOMÉRIEUX INTERNAL INVESTIGATION WAS PERFORMED FOR THE REPORTED FALSE SUSCEPTIBLE ERTAPENEM (ETP) RESULTS FOR TWO KLEBSIELLA PNEUMONIAE STRAINS FROM ONE PATIENT, IN ASSOCIATION WITH THE VITEK® 2 AST-GN70 TEST KIT (LOT 5900705403). THE CUSTOMER'S TWO SUBMITTED STRAINS WERE SUBCULTURED, AND IDENTIFICATIONS OF KLEBSIELLA PNEUMONIAE WERE CONFIRMED. TESTING INCLUDED AST-GN70 CARDS FROM THE CUSTOMER LOT (5900705403) AND A RANDOM LOT (5900650403) ,AS WELL AS AGAR DILUTION (AD) AS THE REFERENCE METHOD FOR THE ETP01N FORMULATION USED IN THIS CARD. ETP BREAKPOINTS: </=0.5 (S) ,1 (I), >/=2 (R). STRAIN 911676 (K. PNEUMONIAE #1): FOR ONE CARD ETP MICS = 1 WITH AES MODIFICATION TO RESISTANT DUE TO IMPERMEABILITY CARBAPENEMASE PHENOTYPE DETECTED. THIS CARD IS AN ESSENTIAL AGREEMENT ERROR WITH THE REFERENCE WITH A MINOR ERROR OF CATEGORY. FOR THE SECOND CARD ETP MIC = 2. ETP (SAME PHENOTYPE DETECTED). ETP REFERENCE AD MIC = 4. THIS CARD IS IN ESSENTIAL AGREEMENT WITH THE REFERENCE METHOD. STRAIN 911677 (K. PNEUMONIAE #2): BOTH CARDS TESTED RESULTED IN ETP MIC = 2 WITH THE SAME IMPERMEABILITY CARBAPENEMASE PHENOTYPE DETECTED. ETP REFERENCE AD MIC = 4. VITEK 2 CARDS AND REFERENCE METHOD ARE IN ESSENTIAL AGREEMENT. THE CUSTOMER'S REPORTED FALSE SUSCEPTIBILITY WAS NOT DUPLICATED. VITEK 2 CARDS ARE PERFORMING AS EXPECTED.
A CUSTOMER IN THE UNITED STATES REPORTED FALSE SUSCEPTIBLE ERTAPENEM RESULTS FOR TWO KLEBSIELLA PNEUMONIAE STRAINS FROM ONE PATIENT, IN ASSOCIATION WITH THE VITEK® 2 AST-GN70 TEST KIT (LOT 5900705403). THE CUSTOMER REPORTED THAT THE ERTAPENEM WAS SUSCEPTIBLE WITH INITIAL TESTING (MIC= <=0.5). UPON REPEAT TESTING ERTAPENEM TESTED RESISTANT (MIC=2 WITH NEW LOWERED CARBAPENEM BREAKPOINTS). ERTAPENEM DISC TESTING WAS RESISTANT WITH ZONE SIZE OF 15 MM. THERE IS NO INDICATION OR REPORT FROM THE HOSPITAL OR TREATING PHYSICIAN TO BIOMÉRIEUX THAT THE DISCREPANT RESULT LED TO ANY ADVERSE EVENT RELATED TO THE PATIENT'S STATE OF HEALTH. A BIOMÉRIEUX INTERNAL INVESTIGATION WILL BE INITIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 573133 | VITEK® 2 AST-GN70 TEST KIT | VITEK® 2 AST-GN70 TEST KIT | LON | BIOMERIEUX, INC | 5900705403 | 03573026398477 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |