FDA Adverse Event Malfunction Summary report: N

CATALYS SYSTEM

MDR report key: 7727218 · Received July 27, 2018

Report

Report Number
3005675890-2018-00061
Event Type
Malfunction
Date Received
July 27, 2018
Date of Event
February 14, 2018
Report Date
November 2, 2020
Manufacturer
JOHNSON & JOHNSON SURGICAL VISION, INC
Product Code
OOE
UDI-DI
05050474609792
PMA / PMN Number
K121091
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE NR-0148919 AND CAPA-010215.

Additional Manufacturer Narrative · 1

JOHNSON & JOHNSON SURGICAL VISION, INC INVESTIGATION SUBSEQUENTLY IDENTIFIED THAT THE CATALYS SYSTEM MAY HAVE A REMOTE POTENTIAL EVENT WHERE LOSS OF SUCTION DURING LASER FIRING COULD CREATE A HAZARDOUS SITUATION THAT MAY RESULT IN SCORING OF THE POSTERIOR CORNEAL SURFACE. THEREFORE JOHNSON & JOHNSON SURGICAL VISION, INC HAS ISSUED AN ADVISORY NOTICE TO REINFORCE INSTRUCTIONS PROVIDED IN PRODUCT TRAINING AND IN THE OPERATOR¿S MANUAL REGARDING SUCTION LOSS DURING A PROCEDURE. A REVIEW OF RECORDS RELATED TO THE DEVICE INCLUDING LABELING, COMPLAINT TRENDING, AND RISK DOCUMENTATION WILL BE PERFORMED. UPON COMPLETION OF THIS REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION OBTAINED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.

Description of Event or Problem · 1

ACCOUNT REPORTED THERE WAS SUCTION LOSS WHILE LASER FIRING. NO ADDITIONAL INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
571086 CATALYS SYSTEM CATALYS OOE JOHNSON & JOHNSON SURGICAL VISION, INC CATALYS-U 05050474609792

Patients

Seq Age Sex Outcome Treatment
1