FDA Adverse Event Injury Summary report: N

AMISTEM-H PROXIMAL COATING STD STEM SIZE 8

MDR report key: 7726662 · Received July 27, 2018

Report

Report Number
3005180920-2018-00571
Event Type
Injury
Date Received
July 27, 2018
Date of Event
July 6, 2018
Report Date
July 27, 2018
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804434
PMA / PMN Number
K161635
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 26 JULY 2018: LOT 171674: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 30 AUGUST 2017. EXPIRATION DATE: 2022-08-17. NO ANOMALIES FOUND RELATED TO THE ISSUE. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

THE PATIENT CAME IN COMPLAINING OF PAIN 1 MONTH AFTER PRIMARY. THE SURGEON DETERMINED THAT THE PATIENT FRACTURED NEAR THE DISTAL END OF THE STEM. THE SURGEON REVISED THE STEM AND HEAD WITH ANOTHER COMPANY'S PRODUCT. THE CUP AND LINER REMAIN IN PLACE. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
567696 AMISTEM-H PROXIMAL COATING STD STEM SIZE 8 CEMENTLESS HIP STEM LZO MEDACTA INTERNATIONAL SA 171674 07630030804434

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention