BD VACUTAINER® BLOOD TRANSFER DEVICE
Report
- Report Number
- 2618282-2018-00037
- Event Type
- Malfunction
- Date Received
- July 26, 2018
- Date of Event
- July 3, 2018
- Report Date
- August 16, 2018
- Manufacturer
- BD CARIBE LTD.
- Product Code
- JKA
- PMA / PMN Number
- K991088
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
CORRECTION: DUE TO AN IT ISSUE BEGINNING ON 7/3/2018, PREVIOUSLY FILED EMDRS DID NOT CONTAIN REQUIRED FIELDS. THIS SUPPLEMENTAL EMDR IS FILED TO PROVIDE THE FOLLOWING OMITTED FIELDS: EVENT ATTRIBUTED TO: OTHER. DEVICE SINGLE USE?: NO. DEVICE RETURNED TO MANUFACTURER: YES.
BD RECEIVED SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE SAMPLES WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR A CLOGGED TRANSFER DEVICE WITH THE INCIDENT LOT WAS NOT OBSERVED. ADDITIONALLY, RETENTION SAMPLES WERE SELECTED FROM BD INVENTORY FOR EVALUATION/TESTING AND UPON COMPLETION, NO ISSUES WERE OBSERVED RELATING TO CLOGGED AS ALL SAMPLES MET SPECIFICATIONS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BASED ON EVALUATION OF THE CUSTOMER AND RETAIN SAMPLES, THE CUSTOMER¿S INDICATED FAILURE MODE FOR A CLOGGED TRANSFER DEVICE WITH THE INCIDENT LOT WAS NOT OBSERVED AS ALL SAMPLES MET THE REQUIRED SPECIFICATIONS. BASED ON THE INVESTIGATION, A ROOT CAUSE COULD NOT BE DETERMINED. THE PRODUCT WAS FOUND TO BE IN CONFORMANCE AND MEET RELEASE SPECIFICATIONS. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. THE BD BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT BD VACUTAINER® BLOOD TRANSFER DEVICE WAS CLOGGED. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.
IT WAS REPORTED THAT BD VACUTAINER BLOOD TRANSFER DEVICE WAS CLOGGED. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.
MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 7102626, MEDICAL DEVICE EXPIRATION DATE: 03/31/2020, DEVICE MANUFACTURE DATE: 05/19/2017. MEDICAL DEVICE LOT #: 8061955, MEDICAL DEVICE EXPIRATION DATE: 02/28/2021, DEVICE MANUFACTURE DATE: 04/12/2018. (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BD VACUTAINER® BLOOD TRANSFER DEVICE WAS CLOGGED. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.