TAXUS LIBERTE PACLITAXEL-ELUTING CORONARY STENT
Report
- Report Number
- 6000089-2006-01668
- Event Type
- Malfunction
- Date Received
- August 11, 2006
- Date of Event
- May 25, 2006
- Report Date
- May 25, 2006
- Manufacturer
- BOSTON SCIENTIFIC
- Product Code
- NIQ
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE ANALYSIS CONFIRMED THE COMPLAINT REPORTED. THE DEVICE APPEARED TO HAVE BEEN KINKED AT THE POINT OF SEPARATION. VISUAL AND MICROSCOPIC EXAMINATION OF THE DEVICE REVEALED THAT THE HYPO-TUBE HAD BROKEN APPROX 62 CENTIMETERS DISTAL FROM THE CATHETER'S STRAIN RELIEF. A REVIEW OF THE MFG RECORDS FOR THIS PARTICULAR BATCH NAMELY TOP ASSEMBLY BATCH # 8233949 FOUND THAT THE DEVICE MET ITS MATERIAL, ASSEMBLY AND PRODUCT SPECS, AT THE TIME OF RELEASE TO DISTRIBUTION. A DETAILED EXAMINATION OF THE RETURNED DEVICE COULD NOT IDENTIFY ANY ISSUE WITH THE ACTUAL DEVICE THAT COULD CAUSE SUCH A BREAK TO OCCUR. THE ROOT CAUSE OF THE COMPLAINT INCIDENT COULD NOT BE DETERMINED.
THIS COMPLAINT IS NOW REPORTABLE BASED ON THE ANALYSIS COMPLETED ON 7/20/06. IT WAS REPORTED THAT DURING A CORONARY DRUG ELUTING STENTING TREATMENT PROCEDURE, CROSSING INABILITY OCCURRED. THE PHYSICIAN ATTEMPTED TO PLACE A TAXUS LIBERTE 2.5X16MM DRUG ELUTING STENT TO THE 85% STENOSED LESION LOCATED IN THE MODERATELY CALCIFIED, SEVERELY TORTUOUS RIGHT CORONARY ARTERY (RCA), BUT WAS UNABLE TO CROSS THE LESION. UPON WITHDRAWAL OF THE STENT, IT WAS NOTED THAT THE SHAFT HAD KINKED DURING THE PROCEDURE. THE PHYSICIAN THEN COMPLETED THE PROCEDURE WITH ANOTHER TAXUS LIBERTE STENT, SAME SIZE. NO PT INJURIES OR COMPLICATIONS OCCURRED. HOWEVER, THE RETURNED PRODUCT CONFIRMED THAT A SHAFT FRACTURE OCCURRED. THIS PRODUCT IS ONLY OUS APPROVED, BUT IT IS SIMILAR TO A MARKETED US DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TAXUS LIBERTE PACLITAXEL-ELUTING CORONARY STENT | DRUG COATED STENT | NIQ | BOSTON SCIENTIFIC | 2.5X16MM | 8233949 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |