SKYTRON
Report
- Report Number
- 1825014-2006-00008
- Event Type
- Other
- Date Received
- August 10, 2006
- Date of Event
- July 26, 2006
- Report Date
- August 9, 2006
- Manufacturer
- KMW GROUP, SKYTRON DIV
- Product Code
- BWN
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
IN 2006, VICTORY PRODUCTS RECEIVED A CALL FROM HOSP AND THE REPORTED SMOKE EMANATING FROM THE POWER CORD OF A 6500 TABLE. THE SKYTRON REP PROCEEDED TO THE FACILITY TO PERFORM AN EVALUATION OF THE SITUATION. HIS PRELIMINARY EXAMINATION REVEALED A BURNT OR MELTED POWER CORD ASSEMBLY. THE CAUSE OF THE BURNT POWER CORD APPEARS TO HAVE BEEN INDUCED BY OXIDATION OR CORROSION OF THE TABLE BASE POWER CORD RECEPTABLE CONNECTOR PINS. THE INCREASED RESISTANCE LEAD TO EXCESSIVE HEAT AND SUBSEQUENT MELTING OF THE FEMALE END OF THE POWER CORD. DURING HIS INVESTIGATION, HE LEARNED THAT THE CLEANING PRACTICES OF THE FACILITY LIKELY CAUSED THE CORROSION TO THE CONNECTOR. THE CLEANING STAFF SPRAY CLEANING FLUID DIRECTLY ONTO THE BASE OF THE TABLE AND CONNECTORS AGAINST THE MFRS MAINTENANCE AND USAGE INSTRUCTIONS. ROUTINE PREVENTATIVE MAINTENANCE BY IN HOUSE PERSONNEL DID NOT DETECT THE CORROSION OR REPLACE THE PARTS. ADDITIONALLY, SKYTRON REQUIRES THAT WHEN USING SPRAY CLEANERS DO NOT SPRAY FLUIDS DIRECTLY INTO A ELECTRICAL CONNECTORS OR MICRO SWITCHES. THIS INFO IS CLEARLY STATED IN THE PRODUCT'S OPERATORS MANUAL. SKYTRON IS WORKING WITH VICTORY PRODUCTS AND ALEXANDRIA HOSP TO RETROFIT THE POWER CORD TO A SEALED TYPE NON DETACHABLE STYLE DUE TO THEIR UNWILLINGNESS TO CHANGE CLEANING PRACTICES. WE ARE STILL IN THIS PROCESS AND WE WILL FILE A FOLLOW UP REPORT WHEN COMPLETE WITH CONCLUSIVE ACTION.
EXCESSIVE SMOKE EMANATING FROM 120 VOLT POWER CORD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SKYTRON | SURGICAL TABLE | BWN | KMW GROUP, SKYTRON DIV | ELITE 6500 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |