AMISTEM COLLARED STEM COLLARED HA COATED LAT STEM SIZE2

MDR report key: 7718470 · Received July 25, 2018

Report

Report Number
3005180920-2018-00544
Event Type
Injury
Date Received
July 25, 2018
Date of Event
June 26, 2018
Report Date
July 25, 2018
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804670
PMA / PMN Number
K121011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 24 JULY 2018: LOT 133015:30 ITEMS MANUFACTURED AND RELEASED ON 02 SEPTEMBER 2013. EXPIRATION DATE: 2018-07-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, 26 ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS MANAGER ON JUL 12, 2018: PARTIAL HIP REVISION SURGERY (STEM, HEAD AND LINER) OCCURRED 3 YEARS AND 6 MONTHS AFTER PRIMARY CEMENTLESS TOTAL HIP ARTHROPLASTY. RADIOGRAPHIC IMAGES PROVIDED SHOW THE PRESENCE OF RADIOLUCENT LINES AROUND THE STEM SUGGESTING MOBILIZATION. ASEPTIC LOOSENING IS A POSSIBLE, LITERATURE DESCRIBED ADVERSE EVENT AFTER CEMENTLESS HIP ARTHROPLASTY AND CAUSES ARE OFTEN UNKNOWN. THE REASON OF THIS FAILURE CANNOT BE DETERMINED.

Description of Event or Problem · 1

THE PATIENT CAME IN COMPLAINING OF PAIN 3 YEARS AND 6 MONTHS AFTER PRIMARY. THE SURGEON DETERMINED THAT THE STEM HAD POTTED. THE SURGEON REVISED THE STEM, HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
561770 AMISTEM COLLARED STEM COLLARED HA COATED LAT STEM SIZE2 CEMENTLESS STEM LZO MEDACTA INTERNATIONAL SA 01.18.242 133015 07630030804670

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention