FDA Adverse Event Injury Summary report: N

AMISTEM H HA COATED LAT STEM SIZE 3

MDR report key: 7718468 · Received July 25, 2018

Report

Report Number
3005180920-2018-00534
Event Type
Injury
Date Received
July 25, 2018
Date of Event
June 25, 2018
Report Date
July 25, 2018
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804199
PMA / PMN Number
K093944
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 24 JULY 2018. LOT 133082: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 07 OCTOBER 2013. EXPIRATION DATE: 2018-08-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. CLINICAL EVALUATION PERFORMED BY MEDICAL AFFAIRS DIRECTOR ON 12 JULY 2018. TOTAL HIP REVISION SURGERY OCCURRED 4 YEARS AFTER IMPLANTATION OF A CEMENTLESS THA IN A (B)(6) YEAR OLD MAN. RADIOGRAPHIC IMAGE PROVIDED SHOWS THE PRESENCE OF SIGNS OF STRESS-SHIELDING. ASEPTIC LOOSENING IS A POSSIBLE LITERATURE DESCRIBED ADVERSE EVENT AFTER CEMENTLESS TOTAL HIP ARTHROPLASTY AND CAUSES ARE OFTEN UNKNOWN. ON THE BASIS OF INFORMATION RECEIVED, THE REASON OF THIS FAILURE CANNOT BE DETERMINED.

Description of Event or Problem · 1

ABOUT 4 YEARS AFTER SURGERY, THE SURGEON REVISED THE PATIENT FOR STEM LOOSENING AND A STRONG INFECTION SUSPICION. THE SURGERY WAS COMPLETED SUCCESSFULLY. ON 12 JULY 2018 WE WERE INFORMED THAT NO INFECTION WAS DETECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
561768 AMISTEM H HA COATED LAT STEM SIZE 3 CEMENTLESS HIP STEM LZO MEDACTA INTERNATIONAL SA 133082 07630030804199

Patients

Seq Age Sex Outcome Treatment
1 76 YR Required Intervention