HANDPIECE MICS
Report
- Report Number
- 3005985723-2018-00443
- Event Type
- Malfunction
- Date Received
- July 25, 2018
- Date of Event
- July 23, 2018
- Report Date
- August 30, 2018
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- UDI-DI
- 00848486030193
- PMA / PMN Number
- K142530
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- OTHER
Narratives
REPORTED EVENT: IT WAS REPORTED THAT THE MICS WAS NON FUNCTIONAL DURING THE CASE LEADING TO A SURGICAL DELAY OF 16-30 MINS PRODUCT EVALUATION AND RESULTS: "VISUAL INSPECTION: VISUAL INSPECTION REVEALED NO PHYSICAL DAMAGE. DIMENSIONAL INSPECTION: DIMENSIONAL INSPECTION WAS NOT COMPLETED. THE REPORTED PROBLEM WAS A FUNCTIONAL ISSUE. MATERIAL ANALYSIS: MATERIAL ANALYSIS WAS NOT COMPLETED BECAUSE THE FAILURE WAS FUNCTIONAL. FUNCTIONAL INSPECTION: FUNCTIONAL INSPECTION REVEALED NO PHYSICAL DAMAGE AND THE PRODUCT FUNCTIONED AS INTENDED. THE HANDPIECE WAS TESTED IN THE HANDPIECE TEST AND PASSED. SEE ATTACHED SCREENSHOT." PRODUCT HISTORY REVIEW: DEVICE HISTORY RECORDS INDICATE 25 DEVICES WERE MANUFACTURED UNDER LOT K09ZT AND 23 WERE ACCEPTED INTO FINAL STOCK ON (B)(6) 2017. A REVIEW OF QT17-08-0042 REVEALED THAT THE ISSUE IS NOT RELATED TO THE FAILURE ALLEGED IN THIS COMPLIANT. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN PRODEX LOT K09ZT SHOWS 2 ADDITIONAL COMPLAINT(S) RELATED TO THE FAILURE IN THIS INVESTIGATION. THE COMPLAINTS ARE 1769269 AND 1787006. CONCLUSIONS: THE ALLEGED FAILURE MODE WAS NOT CONFIRMED. CORRECTIVE ACTION/PREVENTIVE ACTION: A REVIEW OF STRYKER¿S NC/CAPA DATABASE INDICATED THERE HAVE BEEN AN NC AND CAPA ASSOCIATED WITH THE PRODUCT AND FAILURE MODE REPORTED IN THIS EVENT. THIS IS NC1414517 AND CAPA 1450904.
DURING BONE PREP, FOUR OF THE SIX CUTS WERE MADE SUCCESSFULLY. ON THE FIFTH CUT, WHICH WAS THE DISTAL FEMUR, WE FINISHED THE MEDIAL SIDE AND AS WE WENT TO CUT THE LATERAL SIDE, THE MICS LOST POWER. AN ERROR MESSAGE CAME UP WITH THE MESSAGE "FAULT VBUS LOW." THE COMPUTER SUGGESTED RESETTING THE CUTTER, WHICH WE DID. IT SUGGESTED UNPLUGGING THE MICS AND PLUGGING IT BACK IN, WHICH WE DID. WHEN FOLLOWING THESE STEPS, THE ERROR MESSAGE WOULD DISAPPEAR, THEN REAPPEAR A FEW SECONDS LATER. WE FINALLY SHUT THE ROBOT DOWN, RESTARTED IT, HOMED IT, AND ATTEMPTED A MICS STATUS CHECK. THE MICS PRODUCED NO POWER AND GAVE THE LOW BUS VOLTAGE FAULT ERROR AGAIN. WE THEN SHUT THE ROBOT DOWN TWO MORE TIMES, REPLACING THE MICS EACH TIME. WE ATTEMPTED THREE DIFFERENT MICS THAT ALL FAILED BEFORE THE SURGEON DECIDED TO PROCEED MANUALLY TO FINISH THE CASE. AT NO TIME WAS A BURR OVERRIDE ATTEMPTED. AS DR. NIX FINISHED THE CASE MANUALLY, I THEN SHUT THE ROBOT DOWN TWO MORE TIMES AND ATTEMPTED A MICS STATUS CHECK ON EACH OF THE THREE MICS. ALL THREE PASSED AT THIS POINT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT AS DR. NIX FINISHED THE CASE MANUALLY AND UTILIZED THE ROBOT TO CHECK LIGAMENT BALANCING AT THE END OF THE CASE. MICS - 42070617 / 4202483 THIS WAS THE SECOND MICS THAT WE ATTEMPTED TO USE IN DR. NIX'S TKA WHERE THE MICS LOST POWER. WE HAD SHUT THE ROBOT DOWN, RESTARTED IT, HOMED IT, AND ATTACHED THIS MICS TO THE ROBOT. WE ATTEMPTED A MICS STATUS CHECK, AND THE MICS PRODUCED NO POWER. IT GAVE A LOW BUS VOLTAGE FAULT ERROR. MICS - 42050518 / 4204116 THIS WAS THE THIRD MICS THAT WE ATTEMPTED TO USE TO COMPLETE DR. NIX'S ROBOTIC TKA. AFTER THE FIRST TWO MICS HAD FAILED TO WORK, WE SHUT DOWN THE ROBOT AGAIN, RESTARTED IT, HOMED IT, AND ATTACHED THE MICS. WE PERFORMED A MICS STATUS CHECK AND THE MICS PRODUCED NO POWER AND GAVE US THE ERROR MESSAGE "LOW BUS VOLTAGE FAULT." MICS - 42060717 / 4202641 CASE TYPE: TKA WAS THERE A SURGICAL DELAY? YES -16-30 MINUTES.
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
DURING BONE PREP, FOUR OF THE SIX CUTS WERE MADE SUCCESSFULLY. ON THE FIFTH CUT, WHICH WAS THE DISTAL FEMUR, WE FINISHED THE MEDIAL SIDE AND AS WE WENT TO CUT THE LATERAL SIDE, THE MICS LOST POWER. AN ERROR MESSAGE CAME UP WITH THE MESSAGE "FAULT VBUS LOW." THE COMPUTER SUGGESTED RESETTING THE CUTTER, WHICH WE DID. IT SUGGESTED UNPLUGGING THE MICS AND PLUGGING IT BACK IN, WHICH WE DID. WHEN FOLLOWING THESE STEPS, THE ERROR MESSAGE WOULD DISAPPEAR, THEN REAPPEAR A FEW SECONDS LATER. WE FINALLY SHUT THE ROBOT DOWN, RESTARTED IT, HOMED IT, AND ATTEMPTED A MICS STATUS CHECK. THE MICS PRODUCED NO POWER AND GAVE THE LOW BUS VOLTAGE FAULT ERROR AGAIN. WE THEN SHUT THE ROBOT DOWN TWO MORE TIMES, REPLACING THE MICS EACH TIME. WE ATTEMPTED THREE DIFFERENT MICS THAT ALL FAILED BEFORE THE SURGEON DECIDED TO PROCEED MANUALLY TO FINISH THE CASE. AT NO TIME WAS A BURR OVERRIDE ATTEMPTED. AS DR. (B)(6) FINISHED THE CASE MANUALLY, I THEN SHUT THE ROBOT DOWN TWO MORE TIMES AND ATTEMPTED A MICS STATUS CHECK ON EACH OF THE THREE MICS. ALL THREE PASSED AT THIS POINT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT AS DR. (B)(6) FINISHED THE CASE MANUALLY AND UTILIZED THE ROBOT TO CHECK LIGAMENT BALANCING AT THE END OF THE CASE. MICS - 42070617 / 4202483. THIS WAS THE SECOND MICS THAT WE ATTEMPTED TO USE IN DR. (B)(6)'S TKA WHERE THE MICS LOST POWER. WE HAD SHUT THE ROBOT DOWN, RESTARTED IT, HOMED IT, AND ATTACHED THIS MICS TO THE ROBOT. WE ATTEMPTED A MICS STATUS CHECK, AND THE MICS PRODUCED NO POWER. IT GAVE A LOW BUS VOLTAGE FAULT ERROR. MICS - 42050518 / 4204116. THIS WAS THE THIRD MICS THAT WE ATTEMPTED TO USE TO COMPLETE DR. (B)(6)'S ROBOTIC TKA. AFTER THE FIRST TWO MICS HAD FAILED TO WORK, WE SHUT DOWN THE ROBOT AGAIN, RESTARTED IT, HOMED IT, AND ATTACHED THE MICS. WE PERFORMED A MICS STATUS CHECK AND THE MICS PRODUCED NO POWER AND GAVE US THE ERROR MESSAGE "LOW BUS VOLTAGE FAULT." MICS - 42060717 / 4202641. CASE TYPE: TKA. WAS THERE A SURGICAL DELAY? YES -16-30 MINUTES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 562382 | HANDPIECE MICS | STEREOTAXIC DEVICE, ROBOTICS | OLO | MAKO SURGICAL CORP. | 420607174202641 | 00848486030193 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |