TITAN TOUCH SCRO ZERO ANG 20CM
Report
- Report Number
- 2125050-2018-00544
- Event Type
- Injury
- Date Received
- July 24, 2018
- Date of Event
- June 21, 2018
- Report Date
- August 2, 2018
- Manufacturer
- COLOPLAST A/S
- Product Code
- FHW
- UDI-DI
- 05708932539197
- PMA / PMN Number
- P000006
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS FOLLOW-UP MDR IS CREATED TO DOCUMENT THE CONCLUSION OF THE INVESTIGATION. REQUESTS FOR THE EXPLANTED PROSTHESIS AND FOR ADDITIONAL INFORMATION SURROUNDING THIS EVENT HAVE BEEN MADE. HOWEVER, TO DATE NEITHER THE DEVICE NOR THE INFORMATION HAS BEEN RECEIVED. WITHOUT THE BENEFIT OF ANALYZING THE EXPLANT AND WITHOUT THE REQUESTED INFORMATION, QUALITY CANNOT COMMENT ON THE CONDITION OF THE DEVICE OR THE EVENTS SURROUNDING THIS INCIDENT. IF THE EXPLANTED DEVICE OR ADDITIONAL INFORMATION IS RECEIVED, QUALITY WILL RE-EVALUATE THIS EVENT IN ACCORDANCE WITH PROCEDURES. QUALITY REVIEWED THE MANUFACTURING RECORDS FOR THIS DEVICE AND CONFIRMED THAT THERE WERE NO DISCREPANCIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT AND VERIFIED THAT THE DEVICES FROM THIS LOT MET ALL SPECIFICATIONS PRIOR TO RELEASE. A REVIEW OF THE COMPLAINT HISTORY DATABASE REVEALED NO ADDITIONAL COMPLAINTS FOR LOT 4241684. REVIEW OF NONCONFORMING REPORTS REVEALED NO NONCONFORMANCE'S FOR THIS LOT. NO CAPAS ARE ASSOCIATED WITH THIS LOT. MANAGEMENT ROUTINELY REVIEWS EVENTS SUCH AS THIS AND MONITORS COMPLAINT LEVELS. ADDITIONALLY, EVENTS OF THIS TYPE ARE CAPTURED IN THE PRODUCT RISK DOCUMENTATION. BASED ON THIS INFORMATION NO FURTHER CORRECTIVE ACTION IS REQUIRED AT THIS TIME.
WITHOUT THE BENEFIT OF EXAMINATION AND TESTING, COLOPLAST IS PRECLUDED FROM COMMENTING ON THE CONDITION OF THE DEVICE OR THE CAUSE OF THE OCCURRENCE. ONCE OUR EVALUATION IS COMPLETE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
ACCORDING TO THE AVAILABLE INFORMATION, MALFUNCTION - PUMP COLLAPSE, R/O LEAK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 555896 | TITAN TOUCH SCRO ZERO ANG 20CM | INFLATABLE PENILE PROSTHESIS | FHW | COLOPLAST A/S | ES29202400 | 4241684 | 05708932539197 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Other |