FDA Adverse Event Injury Summary report: N

TITAN TOUCH INFR ZERO ANG 18CM

MDR report key: 7712164 · Received July 23, 2018

Report

Report Number
2125050-2018-00540
Event Type
Injury
Date Received
July 23, 2018
Date of Event
June 28, 2018
Report Date
August 22, 2018
Manufacturer
COLOPLAST A/S
Product Code
FHW
UDI-DI
05708932539128
PMA / PMN Number
P000006
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

EXAMINATION OF PHOTOGRAPHS OF THE DEVICE REVEALED A SEPARATION IN THE EXHAUST TUBE OF A CYLINDER NEAR THE STRAIN RELIEF. HOWEVER, QUALITY WAS UNABLE TO DETERMINE WHEN THE SEPARATION MOST LIKELY OCCURRED. THE MOST LIKELY ROOT CAUSE OF THE EVENT CANNOT BE DETERMINED AT THIS TIME. IF THE DEVICE OR ADDITIONAL INFORMATION IS RECEIVED, QUALITY WILL RE-EVALUATE THIS EVENT IN ACCORDANCE WITH PROCEDURES. QUALITY REVIEWED THE MANUFACTURING RECORDS FOR THIS DEVICE AND CONFIRMED THAT THERE WERE NO DISCREPANCIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT AND VERIFIED THAT THE DEVICES FROM THIS LOT MET ALL SPECIFICATIONS PRIOR TO RELEASE. REVIEW OF NONCONFORMING REPORTS REVEALED NO NONCONFORMANCE WITH THIS LOT THAT WOULD HAVE CONTRIBUTED TO THE EVENT. A REVIEW OF THE COMPLAINT DATABASE REVEALED NO SIGNIFICANT TRENDS IN COMPLAINTS OF THIS TYPE FOR LOT 3912598. MANAGEMENT ROUTINELY REVIEWS EVENTS SUCH AS THIS AND MONITORS COMPLAINT LEVELS. ADDITIONALLY, EVENTS OF THIS TYPE ARE CAPTURED IN THE PRODUCT RISK DOCUMENTATION. BASED ON THIS INFORMATION, NO FURTHER CORRECTIVE ACTION IS REQUIRED AT THIS TIME.

Additional Manufacturer Narrative · 1

WITHOUT THE BENEFIT OF EXAMINATION AND TESTING, COLOPLAST IS PRECLUDED FROM COMMENTING ON THE CONDITION OF THE DEVICE OR THE CAUSE OF THE OCCURRENCE. ONCE OUR EVALUATION IS COMPLETE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED DUE TO A TUBING BREAK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
553722 TITAN TOUCH INFR ZERO ANG 18CM INFLATABLE PENILE PROSTHESIS FHW COLOPLAST A/S ES28182400 3912598 05708932539128

Patients

Seq Age Sex Outcome Treatment
1 73 YR Other