TITAN TOUCH INFR ZERO ANG 18CM
Report
- Report Number
- 2125050-2018-00540
- Event Type
- Injury
- Date Received
- July 23, 2018
- Date of Event
- June 28, 2018
- Report Date
- August 22, 2018
- Manufacturer
- COLOPLAST A/S
- Product Code
- FHW
- UDI-DI
- 05708932539128
- PMA / PMN Number
- P000006
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
EXAMINATION OF PHOTOGRAPHS OF THE DEVICE REVEALED A SEPARATION IN THE EXHAUST TUBE OF A CYLINDER NEAR THE STRAIN RELIEF. HOWEVER, QUALITY WAS UNABLE TO DETERMINE WHEN THE SEPARATION MOST LIKELY OCCURRED. THE MOST LIKELY ROOT CAUSE OF THE EVENT CANNOT BE DETERMINED AT THIS TIME. IF THE DEVICE OR ADDITIONAL INFORMATION IS RECEIVED, QUALITY WILL RE-EVALUATE THIS EVENT IN ACCORDANCE WITH PROCEDURES. QUALITY REVIEWED THE MANUFACTURING RECORDS FOR THIS DEVICE AND CONFIRMED THAT THERE WERE NO DISCREPANCIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT AND VERIFIED THAT THE DEVICES FROM THIS LOT MET ALL SPECIFICATIONS PRIOR TO RELEASE. REVIEW OF NONCONFORMING REPORTS REVEALED NO NONCONFORMANCE WITH THIS LOT THAT WOULD HAVE CONTRIBUTED TO THE EVENT. A REVIEW OF THE COMPLAINT DATABASE REVEALED NO SIGNIFICANT TRENDS IN COMPLAINTS OF THIS TYPE FOR LOT 3912598. MANAGEMENT ROUTINELY REVIEWS EVENTS SUCH AS THIS AND MONITORS COMPLAINT LEVELS. ADDITIONALLY, EVENTS OF THIS TYPE ARE CAPTURED IN THE PRODUCT RISK DOCUMENTATION. BASED ON THIS INFORMATION, NO FURTHER CORRECTIVE ACTION IS REQUIRED AT THIS TIME.
WITHOUT THE BENEFIT OF EXAMINATION AND TESTING, COLOPLAST IS PRECLUDED FROM COMMENTING ON THE CONDITION OF THE DEVICE OR THE CAUSE OF THE OCCURRENCE. ONCE OUR EVALUATION IS COMPLETE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
ACCORDING TO THE AVAILABLE INFORMATION, THE DEVICE WAS EXPLANTED DUE TO A TUBING BREAK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 553722 | TITAN TOUCH INFR ZERO ANG 18CM | INFLATABLE PENILE PROSTHESIS | FHW | COLOPLAST A/S | ES28182400 | 3912598 | 05708932539128 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Other |